9
BP868F / UME-07-REV.2-08B13
PREPARATION
3. Preparation
1) Connecting Power Cord
①
Connect the power cord to the AC power connector of the blood pressure monitor.
②
Connect the grounding cable to the grounding terminal.
[Attention] Do not connect the grounding line of the blood
pressure monitor to gas pipe, telephone line, and water pipe.
It is very dangerous.
2) Replacing Fuse
If it does not display anything after you connect the power cord
to the main unit, plug it to the outlet, and turn on the power,
check the fuse first. Using regular screwdriver, remove the fuse
above the power connector on the back of the blood pressure
monitor, and check the wire inside the fuse is broken. If the wire
is broken, replace it as follows. Inside the black fuse holder,
there is an extra fuse. Remove the broken fuse and replace it with
the new fuse.
Extra Fuse
Fuse in Use
[Fuse Holder]