User Manual for AML Oceanographic’s Micro SVT Sensor
22
>set detect 05
(disable the auto baud and default to 9600 baud at power up)
>set startup noheader
(do not display a header at power up)
>set startup scan
03/05/03 14:23:06:02 000.40 1523.74 11.793
>
Command:
SET STARTUP PROMPT [SE S P]
This command instructs the Micro SVT to display a prompt at power on.
If the auto baud feature has
not
been disabled, the Micro SVT will wait for a
carriage return
before it outputs a prompt. (See SET DETECT command).
The following example instructs the Micro SVT to go directly to a prompt at power up.
Example:
set detect 07
(disable the auto baud feature and set the baud rate to 38400)
set startup noheader (disable the header)
set startup prompt (places the Micro SVT in the prompt mode)
5.1.3.6 Accessing the calibration coefficients
Refer to Section 8 and appendices A and B for more information.
The instruments calibration coefficients are stored in the Micro SVT’s memory.
Accessing the coefficients for viewing or editing can be achieved by talking to the
individual sensor boards using the TALK command. The addresses of the sensor boards
are as follows,
Sensor board
Address
Pressure & Temperature
200
Sound Velocity
203
Analog
202
Once communications is established with the individual sensor board, the user can view
or edit the calibration coefficients by using the DIS COEFFICIENTS or SET commands.
Command:
TALK n [TA n]
This command, used in conjunction with the DIS or SET commands, allows access to the
instruments calibrations coefficients for viewing or editing. “n” is the address of the
sensor board as described below. To exit the
talk mode
, use the CTRL C command.
Sensor board
Address
Pressure & Temperature
200
Sound Velocity
203
Analog
202
Example:
>talk 200 (talk to the pressure and temperature board)