User Manual for AML Oceanographic’s Micro SVT Sensor
19
5.1.3.3 Set scan options
Command:
SET SCAN DELAY nnn [SE SC D nnn]
The
scan delay
is defined as the time between the sensor board power up and a scan of
data. The time (nnn) is entered in 0.2-second intervals. Maximum
scan delay
is 255, or 51
seconds.
This command is used for sensors that have a slower power up response time.
Example:
>set scan delay 20 (sets the scan delay to 4 seconds)
In the above example, the Micro SVT would power up the sensor boards, wait 4
seconds then take a scan of data.
Micro
SVT
power up timing diagram
Command:
SET SCAN TIME [SE SC T]
This command enables the displaying of time during a data scan.
Command:
SET SCAN NOTIME [SE SC NOT]
This command disables the displaying of time during a data scan.
Command:
SET SCAN DATE [SE SC DA]
This command enables the displaying the date during a data scan.
Command:
SET SCAN NODATE [SE SC NOD]
This command disables the displaying of the date during a data scan.