User Manual for AML Oceanographic’s Micro SVT Sensor
13
5.1.1.3 Raw Output Mode
The Raw mode shows the outputs from the analog to digital converters for each sensor.
No compensation or manipulation of the data is performed in Raw mode. Refer to section
8 and appendix B to convert the raw data values to engineering data. The raw counts for Nl,
Nsv, Nh, and Nt are integers between 0 and 65535. The output is sent as space delimited
ASCII characters in the following format:
Date(Month/Day/Year) Time (hh:mm:ss:and hundredths of seconds), Nl, Nsv, Nh, Nt and
a carriage return/line feed (end of scan).
Example scan: 04/14/03 14:23:08.06 15283 22544 61897 13487
5.1.2 Command Summary
All commands are in the form of standard English words. Commands can be entered in
upper or lower case letters followed by an <
ENTER
>. The minimum letters of the
command that the instrument will recognize are enclosed in brackets.
Command:
RAW [R]
This command will set the instrument to output RAW uncorrected data when using the
MONITOR or SCAN commands.
Command:
REAL [RE]
This command will set the instrument to output Real corrected engineering data when using
the MONITOR or SCAN commands.
Command:
/ [/]
This command is used to toggle RAW and REAL modes of operation.
Command:
SCAN [S]
This command outputs one scan of data.
Command:
MONITOR [M]
This command sets the instrument to output multiple scans continuously at the scan rate set
by the “SET” command. To discontinue the monitor command the break key or the space
bar can be depressed. Note: this command is not available with the binary format
instruments.