Temperature
Operation
Storage
-20 to +45°C
-20 to +60°C
-20 to +45°C
-20 to +60°C
Dimensions
Diameter
Length
Weight*
Weight in water*
7.6 cm / 3.0 in
33.4 cm / 13.15 in (SVP set up)
1.36 kg / 3.00 lb
.69 kg / 1.52 lb
7.6 cm / 3.0 in
33.4 cm / 13.15 in (SVP set up)
2.5 kg / 5.6 lb
1.9 kg / 4.1 lb
*With a medium cage and no sensors, deployed weights will vary
1.2 Communicating with the Instrument
The preferred method of communication (setup, downloading data, etc.) is through the use of
WiFi (500m only, for a 6000m use USB-C). By default, the instrument automatically sets up a
WiFi hotspot upon startup, allowing WiFi-enabled computers, laptops, and tablets to directly
connect.
The 500m versions of AML-3 LGRs contain a USB-WiFi dongle. Use of the dongle is not
required (any WiFi adapter will connect to the instrument) but it does provide a convenient
method for allowing a compatible PC to host two WiFi connections - the on-board WiFi adapter
to remains connected with the local network, and the USB adapter to connects directly to the
instrument. AML’s software, Sailfish, allows the user to selectively use available WiFi adapters
to connect to the instrument. If you want to connect over WiFi without Sailfish, the password is
the serial number twice. For example, AML_A30001 has a password of A30001A30001.
Note:
You should never be required to enter the password when connecting with Sailfish.
AML-3 LGR instruments come equipped with a USB-C connector with waterproof cover. This
connector provides power to the instrument when connected to a compatible USB port or wall
wart. This connector is also used as a secondary means for data transfer (downloading
internally logged data) and may be configured via AML’s interface software, Sailfish.
Note: This
connector is only waterproof when the protective cover is installed. It is not intended for
use when deployed. Do not remove the waterproof cover when the instrument is wet.
Applying connector grease to the port can increase resistance to damage from a splash
of water when the cover is off.
AML-3 LGR instruments come with a USB-C to USB-A adapter cable to maximize PC
compatibility. The instrument may be connected to any USB-C port using a standard USB-C
interconnect cable, so if the provided cable is lost it is feasible to purchase an off the shelf
replacement, provided the cable is rated to handle 5W. Generally speaking the shorter the cable
the better.
See
for USB-related questions.
AML Oceanographic Ltd.
8/38