dep (Calculated Depth)
sal (Calculated Salinity)
den (Calculated Density)
sound (Calculated Sound Velocity)
Note: Calculated parameters will output
-99.9999 if outside of the valid range of the
calculation formula.
set scan no<parameter>
Remove a parameter from being included in
the log file and monitor output.
set derive <parameter> [y | n]
Enable/disable a parameter to be calculated if
the required sensors are installed.
Derived Parameters:
depth (from pressure)
salc (salinity from conductivity)
salsv (salinity from sound velocity)
density (requires salinity)
sv (sound velocity, requires salinity)
Note: a parameter will only be output if it is
ALSO included in monitor scans. Refer ‘SET
SCAN’ command, above.
5.4 UV Antifouling
Note that these commands apply only in the instance of a UV system installed on the
instrument.
Command
Description
display uv
Print the current UV settings.
set uv time <ontime> <offtime>
Set the duty cycle of the UV lantern in seconds.
Note: Setting UV time to 0 0 will disable light emission.
5.5 Data Logging
Command
Description
display status
Print the current logging status, including in-water status,
number of samples recorded, log duration, and sample
rate.