7 Appendix C - Calculated Parameters
The AML-3 instrument has the ability to calculate certain parameters depending on your sensor
configuration. Below is a table outlining the parameters that can be calculated.
Parameter
Required Sensors
Equation
Range
Salinity
C, T, P
Practical Salinity Scale 1978
0 - 90 PSU
1
Salinity (from SV)
SV, T, P
AML
0 - 90 PSU
Density
C,T, P
TEOS10
0 - 2000 kg/m
3
Density (from SV)
SV,T, P
TEOS10
0 - 2000 kg/m
3
Sound Velocity*
C,T, P
Chen & Millero
0 - 3000 m/s
Depth
P
UNESCO
-20 - 12000 m
C - conductivity, T- temperature, SV - sound velocity, P- pressure
1 - Instrument will report 0-90 PSU, however PSS-78 equation is valid only from 2-50 PSU.
*Calculated salinity must be enabled for the calculated Sound Velocity to operate correctly
AML Oceanographic Ltd.
32/38