INPUT  CURRENT

        Typically,  0.8  A  rms  with  a

48-W  dc  load and a 120-V  ac  input.  Protected  with
a  2-A  fuse  on  the  100-  and  120-V  ac  settings,  and
by a 1-A fuse on the 220- and 240-V ac settings.

DC OUTPUT   

Maximum  rated output  currents  are: 

DC Voltage

Maximum Current

+12 V

  1 A

!

12 V

  1 A

+24 V

0.5 A

!

24 V

0.5 A

Maximum dc  output power from  0 to  50

E

C  is 48W.

Derate 2.5%/

E

C for 50 to 60

E

C.

DC REGULATION  

 Variations in dc output voltages

are  <±0.1  %  over  the  combined  range  of  zero
through  full  load  and  input  voltages  from  t he
minimum  to  maximum  limits  of  the  allowed  input
volt age  range.  Measurements  are  made  within a  1
-minute period. Regulation <±0.3% over any 24-hour
period  at  constant  ambient  temperature  for  the
same  load  and  input  voltage  ranges    after  a  60-
minute warmup. 

LONG-TERM  STABILI TY   

  DC  output  voltages

change  <±0.5%  (after  a  60-minute  warmup) over a
6-month  period at  constant  load, input  voltage,  and
ambient temperature. 

OUTPUT IMPEDANCE

   <0.3 

Ù

 at any frequency up

to 100 kHz for the dc outputs. 

TEMPERATURE  COEFFICIENT

      <0.02%/

E

C  from

0 to 60

E

C for the dc outputs. 

NOISE  AND RIPPLE

   <3 mV peak-to-peak  for  any

dc  output,  as  obs erved  on  a  50-MHZ  bandwidth
oscilloscope. 

VOLTAGE  ADJUSTMENT

      ±5%  minimum  range.

Range  typically  ±1  V  about  the  nominal  supply
voltage. Resettability <±0.05% of the supply voltage. 

RECOVERY TIME

   <50 

ì

s to return to within ±0.1%

of  the  rated  voltage for  all  dc  out puts  for  any  input
voltage  change  within  the  rated  range,  or  for  a
change of load current from 10  to  100%  of  full  load. 

THERMAL  PROTECTION

      The  red,  front-panel,

TEMP  light  turns  on  when  the  temperature  of  the
side-panel heat sink is within 13

E

C of the maximum

operating  temperature  limit  When  t he  heat  sink
temperature  exceeds  the  95

E

C  maximum  limit,  a

thermal  switch  automatic ally  turns  off  the  power
supply. Under this thermal shut-down condition, both
t he  TEMP  and  the  POWER  lights  are  turned  off.
Recovery from thermal shut-down is automatic once
the  load  is  reduced,  so  that   the  heat  sink
temperature  drops  below  the  maximum  operating
limit.
 

CIRCUIT PROTECTION

   Fuse Rating:  2  A (FAST)

size  3AG  fuse  for  100-V  or 120-V operation; 1 A(F)
size 5×20 mm fuse for 220-V or 240-V operation. All
dc outputs include a current  foldback  circuit to limit
the  out put  c urrent to  nominally  150%  of  the  rated
value.  This  feature  provides  short-circuit  and
overload  protection.  Recovery  is  automatic  after
removal of the overload condition. 

2.2.  CONTROLS AND INDICATORS

POWER

      Front-panel  two-position  roc k er  switch

turns  power  on or  off.  Adjacent  red  light  indic at es
power  On  condition  when  illuminated.   Power
indicator  light  and  output  power  turn  off  if  t he  heat
sink temperature exceeds 95

E

C. 

TEMP

   Front-panel red light turns on when the heat

sink  temperature  exceeds  82

E

C  to  warn  that  the

shutdown  temperature  limit  is  being  approached.
TEMP  light turns off  if  power  has  been  shut  off  by
exceeding the heat sink temperature limit. 

2.3.  INPUTS 

AC POWER

     Rear-panel,  Corcom™, international-

standard  IEC  power  connector,  type  CEE-22,
accepts  power  cables  wired  according  to  local
electrical  standards. A power cable  is  shipped  with
the  4002P.  The  Corcom  CEE-22  connector  meets
standard  22  of  the  International  Commission  on
Rules for the  Approval  of  Electrical  Equipment. The
primary  voltage  selector  and  primary  fuse  are
incorporated  into  the  AC  POWER  connector.  The
primary voltage  selected  (100,  120, 220,  or  240  V)
is  visible  through the rear door. U.S.A.  standard  in-
line  fuse  is  2  A for 100 or 120 V ac and 1 A for  220
or 240 V ac. See Sections 

3.2

 and 

3.3

, respectively,

for instructions  on  changing the  voltage setting and
fuse. 

2.4.  OUTPUTS

PREAMP 1, PREAMP  2,  PREAMP  3,  PREAMP  4

Rear-panel,  9-pin,  "D"  connectors  (Amphenol
17-10090)  provide  power  for  up  to 

four

preamplifiers. Connectors mate with power cords on
all  standard  ORTEC  preamplifiers.  Compatible  with

Содержание ORTEC 4002P

Страница 1: ...Model 4002P Portable Power Supply Operating and Service Manual Printed in U S A ORTEC Part No 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge Tennessee 37830 United Statesof America...

Страница 2: ...that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the f...

Страница 3: ...ACCESSORIES 3 2 7 ORDERING INFORMATION 3 3 INSTALLATION 3 3 1 UNPACKING 3 3 2 SELECTING THE MAINS VOLTAGE 4 3 3 REPLACING THE FUSES 4 3 4 LOCATION 4 3 5 COMPATIBILITY 4 4 OPERATING INSTRUCTIONS 5 4 1...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...display the selected input voltage Also incorporated in this connector block are a fuse holder and the input power cord connector The 4002P uses an international standard IEC power connector to permi...

Страница 8: ...t down condition both the TEMP and the POWER lights are turned off Recovery from thermal shut down is automatic once the load is reduced so that the heat sink temperature drops below the maximum opera...

Страница 9: ...monitor each of the dc voltages delivered to the rear panel connectors 2 5 ELECTRICAL AND MECHANICAL WEIGHT Net 10 9 kg 24 lb Shipping 13 kg 29 lb DIMENSIONS 21 5 cm W 26 cm D 10 2 cm H 8 4 in W 10 2...

Страница 10: ...n the door window and firmly press the drum back into the holder Close the module door You are now ready to reconnect the instrument to the ac power source 3 3 REPLACING THE FUSES The 4002P requires t...

Страница 11: ...e is reached an internal cutout in the power supply removes power and neither lamp will be illuminated Test jacks on the panel allow convenient checking of the power supply voltages from the front pan...

Страница 12: ...egulator D4 which receives its unregulated voltage from a voltage doubler circuit consisting of C1 C2 D1 and D2 Current flow to the output is monitored by sensing the voltage drop across R11 When the...

Страница 13: ...6 3 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensure the same quality control and checkout that ar...

Страница 14: ...8...

Отзывы: