background image

3

Pin Number

Power Voltage

1

 Ground

2

 Ground

3

No connection

4

+12 V

5

No connection

6

-24 V

7

+24 V

8

No connection

9

-12 V

Table 1.  Pin Assignments for Preamp Power

Connectors.

Pin Number 

Power Voltage

16

+12 V

17

!

12 V

28

+24 V

29

!

24 V

34

Power Return Ground

42

High-Quality Ground

Table 2.  Pin Assignments as Wired for the NIM

POWER Connector on the Model 4002P.

preamplifier  power  cables  of  most   other  NIM
manufacturers.  Pin  assignments  are  listed  in
Table 

1

.  Each  preamp  power  output  connector  can

be  expanded  to  three  output  connectors  by  using
the  optional  Model  4002P-C1  Preamp  Power  Fan-
Out Cable. Using four of the Model 4002P-C1 allows
the 4002P to supply power to 12 preamplifiers.

NIM  POWER

      Rear-panel,  NIM  standard  bin

connector  compatible with the power  connector  on
the  rear  of  NIM  modules, per TID-20893  (Rev). For
use with a Module Extender Cable, such as  ORTEC
Model  401-C3,  to  power  a  single  NIM  module.  Pin
assignments  are  listed  in  Table 

2

.  Pins  not  listed

have no connection in the Model 4002P, but may be
assigned to  a  specific  function  by TID-20893  (Rev).

OUTPUT TEST POINTS

   Front-panel jacks provide

t est  points  to  monitor  each  of  the  dc  volt ages

delivered to the rear-panel connectors..

2.5.  ELECTRICAL AND MECHANICAL

WEIGHT

   

Net 10.9 kg (24 lb)
Shipping 13 kg (29 lb)

.

DIMENSIONS   

21.5 cm W × 26 cm D × 10.2 cm H

(8.4 in. W × 10.2 in. D × 4.0 in. H).

2.6.  ACCESSORIES

121-Cl  PREAMPLI FI ER 

POW ER 

CABLE

EXTENDER

   Provides a 3-m (10-ft) extension cable

to  connect  a  preamplifier  power  cable  to  the
PREAMP  1,  2, 3, or  4  power  output  on  the 4002P
Portable Power Supply. Compatible with all standard
ORTEC preamplifiers. 

401-C3 MODULE  EXTENDER  CABLE

      Connects

the  4002P NIM POWER output to  the  standard bin
power  connector  on  the  rear  of  a  NIM  module.
Permits powering a NIM module in a remote location
using  the  4002P  Portable  Power  Supply.  Cable
length is 91 cm (3 ft).  

4002P-C1  PREAMP  POWER  FAN-OUT  CABLE

Plugs  into  a  standard  ORTEC  preamplifier  power
output  connector  and  provides  three  preamplifier
power  output  connectors.  Use  four  of  the  optional
4002P-C1 with the 4002P to power 12 preamplifiers. 

2.7.  ORDERING INFORMATION

To  order  the  4002P  Portable  Power  Supply  and/or
accessories,  specify  the  following  model  numbers
and descriptions. 

Model          Description

 

4002P

Portable Power Supply 

121-01

Preamplifier Power Cable Extender

401 -C3 

Module Extender Cable 

4002P-C1

Preamp Power Fan-Out Cable

3.  INSTALLATION

3.1.  UNPACKING

Unpack  the  unit,  being careful  to  retain  all  packing

Содержание ORTEC 4002P

Страница 1: ...Model 4002P Portable Power Supply Operating and Service Manual Printed in U S A ORTEC Part No 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge Tennessee 37830 United Statesof America...

Страница 2: ...that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the f...

Страница 3: ...ACCESSORIES 3 2 7 ORDERING INFORMATION 3 3 INSTALLATION 3 3 1 UNPACKING 3 3 2 SELECTING THE MAINS VOLTAGE 4 3 3 REPLACING THE FUSES 4 3 4 LOCATION 4 3 5 COMPATIBILITY 4 4 OPERATING INSTRUCTIONS 5 4 1...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...display the selected input voltage Also incorporated in this connector block are a fuse holder and the input power cord connector The 4002P uses an international standard IEC power connector to permi...

Страница 8: ...t down condition both the TEMP and the POWER lights are turned off Recovery from thermal shut down is automatic once the load is reduced so that the heat sink temperature drops below the maximum opera...

Страница 9: ...monitor each of the dc voltages delivered to the rear panel connectors 2 5 ELECTRICAL AND MECHANICAL WEIGHT Net 10 9 kg 24 lb Shipping 13 kg 29 lb DIMENSIONS 21 5 cm W 26 cm D 10 2 cm H 8 4 in W 10 2...

Страница 10: ...n the door window and firmly press the drum back into the holder Close the module door You are now ready to reconnect the instrument to the ac power source 3 3 REPLACING THE FUSES The 4002P requires t...

Страница 11: ...e is reached an internal cutout in the power supply removes power and neither lamp will be illuminated Test jacks on the panel allow convenient checking of the power supply voltages from the front pan...

Страница 12: ...egulator D4 which receives its unregulated voltage from a voltage doubler circuit consisting of C1 C2 D1 and D2 Current flow to the output is monitored by sensing the voltage drop across R11 When the...

Страница 13: ...6 3 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensure the same quality control and checkout that ar...

Страница 14: ...8...

Отзывы: