background image

4

Fig. 1.  Setting the Line Voltage and Replacing the
Fuses.

NOTE

The spindles on each side of the drum are octagonal, not
round. This gives each of the 4 settings a definite detent.
To avoid eroding these detents, always remove the drum
before rotating it.

CAUTION

Do not use makeshift fuses or short-circuit the fuse
holders.

materials  until  the  unit  has  been  checked  for
pos sible  concealed  damage.  The  power  cord  is
packed  with  the  unit   and  attaches  to  a  3-pin
connector that  is  mounted  on the rear of the 4002P
Portable Power Supply

3.2.  SELECTING THE MAINS VOLTAGE

The  4002P  uses  a  Corcom  Power  Entry  Module
(Fig.  1),  whic h  contains  the  power  connector;  a
s mall,  removable  drum  for  setting  the  line  voltage;
and  a  line  fuse.  The  removable  drum  lets  you
configure  the  instrument  for  a  nominal  100,  120,
220,   or  240  V  ac.  The  module  door  has  a  small
window  that  shows  the  line  volt age  setting  now  in
use. 

To change the ac input voltage: 

Disconnect  t he  instrument  from  the  ac  power
source. Use a small  flat-blade screwdriver or similar
tool  to  pop  open  the  module  door.  Remove  the
plastic drum: pinch the sides  of  the  drum  with your
fingernails and pull,  or insert a flat-blade screwdriver
in the slot between the 

120Vac

 and 

240Vac

 settings

and  gently  pry  the drum loose,  one  end  at  a  time.
Rotate the drum until the desired line voltage setting
will  s how in  the  door  window,  and  firmly  press  the
drum back into the holder. Close 

the  module door.   You  are  now  ready  to  reconnect
the instrument to the ac power source.

3.3.  REPLACING THE FUSES

The 4002P requires the following fuses:

!

For  100 or 120 V  ac  line  voltage,  a  2-A  (FAST)
size 3AG fuse.

!

For  220 or 240  V ac  line  voltage,  a 1 A(F)  size
5×20-mm fuse.

The 3AG fuse requires a Corcom LA200 fuseholder.
The  5×200-mm  fuse  requires  a  Corcom  LA201
fuseholder.  (We  supply  the  appropriate  fuse  and
fuseholder based on your order specifications.)

To replace a fuse: 

Disconnect  the  instrument  from  the  ac  power
source. Use a small flat-blade screwdriver or similar
tool to pop open the module door. The line fuse is in
a  plastic  fuseholder  drawer  marked  with  an  arrow
(

). Gently pull the fuseholder out and replace the

fuse.  If  changing  to  the  metric type fuse be sure to
install  the  corresponding  fuse  holder.  Slide  the
fuseholder  back  into  the  module  and  close  the
module  door.  You  are  now  ready  to  reconnect  the
instrument to the ac power source.

3.4.  LOCATION

The  ORTEC  4002P  Portable  Power  Supply  is  a
completely  portable,  self-contained  unit  that  is
designed for table-top installation. The basic  design
of  t he  cabinet  provides  for  cooling  of  the  power
supply by natural convection flow. The power supply 
should  not  be  positioned  so as to  block  the  flow  of
air  around  and  through  the  cabinet.   Avoid  locating
the  power  supply  near  or  above  heat-producing
equipment  without  accounting  for  the  temperature
rise.

3.5.  COMPATIBILITY

The  ORTEC  4002P  Port able  Power  Supply  is

Содержание ORTEC 4002P

Страница 1: ...Model 4002P Portable Power Supply Operating and Service Manual Printed in U S A ORTEC Part No 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge Tennessee 37830 United Statesof America...

Страница 2: ...that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the f...

Страница 3: ...ACCESSORIES 3 2 7 ORDERING INFORMATION 3 3 INSTALLATION 3 3 1 UNPACKING 3 3 2 SELECTING THE MAINS VOLTAGE 4 3 3 REPLACING THE FUSES 4 3 4 LOCATION 4 3 5 COMPATIBILITY 4 4 OPERATING INSTRUCTIONS 5 4 1...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...display the selected input voltage Also incorporated in this connector block are a fuse holder and the input power cord connector The 4002P uses an international standard IEC power connector to permi...

Страница 8: ...t down condition both the TEMP and the POWER lights are turned off Recovery from thermal shut down is automatic once the load is reduced so that the heat sink temperature drops below the maximum opera...

Страница 9: ...monitor each of the dc voltages delivered to the rear panel connectors 2 5 ELECTRICAL AND MECHANICAL WEIGHT Net 10 9 kg 24 lb Shipping 13 kg 29 lb DIMENSIONS 21 5 cm W 26 cm D 10 2 cm H 8 4 in W 10 2...

Страница 10: ...n the door window and firmly press the drum back into the holder Close the module door You are now ready to reconnect the instrument to the ac power source 3 3 REPLACING THE FUSES The 4002P requires t...

Страница 11: ...e is reached an internal cutout in the power supply removes power and neither lamp will be illuminated Test jacks on the panel allow convenient checking of the power supply voltages from the front pan...

Страница 12: ...egulator D4 which receives its unregulated voltage from a voltage doubler circuit consisting of C1 C2 D1 and D2 Current flow to the output is monitored by sensing the voltage drop across R11 When the...

Страница 13: ...6 3 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensure the same quality control and checkout that ar...

Страница 14: ...8...

Отзывы: