7 - 11
Cyclops L
Portable Thermometers
User Guide
7.6 Menu Mode - Icon Descriptions
The information in this chapter should be used in conjunction with the
navigational flow chart (Fig. 7-7).
7.6.1 Exit
When in Menu Mode, clicking on the exit icon will return you from any sub-menu.
7.6.2 Emissivity
Use the
and
keys to enter the required value. For information on
emissivity values, see Section 11 - Emissivity. The default value is 1.00.
*Available when
Operator Emissivity Mode
is selected.
7.6.3a Record Mode
Readings taken in Classic, Burst or Latched mode can be stored within the
Cyclops
for later download. The recording facility can be switched on or off by
selecting the icons (i)
On
or (ii)
Off
.
Readings are stored against current time or date set on the clock (See Section
7.4.13).
The readings are also stored against a 8-character alpha-numeric location ID,
which can be set by selecting the
ID
icon (iii). Having set the location ID, it is
possible to auto-increment the ID on each trigger press, See Section 8.
The stored readings can be downloaded by selecting the
Download
icon (iv).
The data is streamed to both the Bluetooth and USB outputs. The protocol of
the streamed data is given in Appendix 1.
The stored readings can be deleted by selecting the
Delete
icon (v). To prevent
accidental deletion, this icon requires a double entry.
The stored readings can be reviewed by selecting the
Review
icon (vi).
(i)
(ii)
(iii)
(iv)
(v)
(vi)
or
*
*
*
*