background image

Adjustment Manual

Industrial Pressure Calibrator

JOFRA IPI & IPI Mk.II

Содержание IPI Mk.II

Страница 1: ...Adjustment Manual Industrial Pressure Calibrator JOFRA IPI IPI Mk II...

Страница 2: ...om the factory with all setting access unlocked or available to be changed 2 In supervisory mode each of the parameters can be locked or unlocked using the and keys Select LOC lock for those parameter...

Страница 3: ...y setting a factor of 25 a 40 PSI pressure would display as 1000 40 x 25 The engineering unit displayed on the IPI will be Cust IPI Calibration Procedure Overview Calibration adjustment of the IPI is...

Страница 4: ...needed to connect to the pressure standard Always make sure the hose tubing and fittings etc have a rated working pressure at or above the pressure of the unit Also it is important that there be no l...

Страница 5: ...put 437 50 psi 87 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below...

Страница 6: ...i 50 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressu...

Страница 7: ...psi 12 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the P...

Страница 8: ...PC such as Hyperterminal or PcPlus Connect the RS232 cable to stereo jack on the back of the IPI The other end of the cable should be connected to the terminal PC serial port An adapter may be needed...

Страница 9: ...enu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Second point above ambient Temp Comp 7 First point below ambi...

Страница 10: ...using the serial port use the VAL command Calibration Menu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Secon...

Страница 11: ...94757 34 473785 35 00 6 894757 3447 3785 3500 437 50 0 06894757 30 164562 30 00 6 894757 3016 4562 3000 375 00 0 06894757 25 855339 26 00 6 894757 2585 5339 2600 312 50 0 06894757 21 546116 22 00 6 89...

Страница 12: ...ment tasks from handheld field instruments to laboratory reference level bench top instruments JOFRA JF Marine Instruments A complete range of calibration equipment for temperature pressure and signal...

Отзывы: