Procedure
After all connections have been made, type 3 {enter}. Use the
Pressure Standard to input –approximately 200,150,100,50,0,
and -12 psi as prompted. When the value is stable, type in the
actual value at the keyboard. The screen should look as shown
below.
After inputting –12 psi the main calibration menu will reappear.
Notes on Calibration Verification:
This procedure covers calibration only. When verifying a unit,
as a rule of thumb the unit should never exceed 50% of its spec.
For verification using the serial port use the VAL? command.
Calibration Menu
1: Calibrate Temperature
2: Set Sensor Parameters
3: Calibrate Pressure
4: Reset Temp Compensation
5: First point above ambient Temp Comp
6: Second point above ambient Temp Comp
7: First point below ambient Temp Comp
8: Second point below ambient Temp Comp
9: Exit and save Data
10: Exit WITHOUT saving Data
Input ~200.000000 psi, Enter the actual value when stable:
Input ~150.000000 psi, Enter the actual value when stable:
Input ~100.000000 psi, Enter the actual value when stable:
Input ~50.000000 psi, Enter the actual value when stable:
Input ~0.000000 psi, Enter the actual value when stable:
Input ~-12.000000 psi, Enter the actual value when stable: