Use the Pressure Standard to output 312.50 psi (62.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show -------. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 250.00 psi (50%). After the output
has stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show -------. When the readings are complete
the screen should look as shown below.
Use the Pressure Standard to output 187.50 psi (37.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show -------. When the readings are
complete the screen should look as shown below.
PSI
250.00
CAL MODE
PSI
187.50
CAL MODE
PSI
125.00
CAL MODE