Use the Pressure Standard to output 125.00 psi (25%). After the output
has stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show -------. When the readings are complete
the screen should look as shown below.
Use the Pressure Standard to output 62.50 psi (12.5%). After the
output has stabilized, press the ENTER key to continue. As the unit
takes readings, the screen will show -------. When the readings are
complete the screen should look as shown below.
Use the Pressure Standard to output 0.00 psi. After the output has
stabilized, press the ENTER key to continue. As the unit takes
readings, the screen will show -------. When the readings are complete
the screen should look as shown below.
PSI
62.50
CAL MODE
PSI
0.00
CAL MODE
PSI
-12.00
CAL MODE