Ametek IPI Mk.II Скачать руководство пользователя страница 4

 

 

 

 

Calibration Interval

 

You should check performance of  the IPI at the interval required by 
your calibration program.  We recommend adjustment when 
measurement deviates by more than 75% of the specified accuracy, or 
0.04% 

 
Test Equipment 

Verification and calibration of the IPI requires pressure and/or vacuum 
standards able to produce and indicate pressures from vacuum to the 
full-scale range of the unit under test.  In order to maintain the specified 
accuracy of the IPI, standards should have a TUR of 4:1 or better. 
 

Connections 

The IPI uses a 1/4” NPT male connection in the pressure input port.  
Various adapters may or may not be needed to connect to the pressure 
standard.  Always make sure the hose, tubing, and fittings etc have a 
rated working pressure at or above the pressure of the unit.  Also it is 
important that there be no leaks when performing calibration; use Teflon 
tape where appropriate.   

 

 

Manual Calibration 

 
Entering Calibration Mode: 

After you have made your connections, turn the power on while holding 
the CONFIG key.  Use the arrow keys to enter the password.  The 
password is 101.  If you have entered calibration mode correctly the 
display should look as shown below.  The pressure value displayed will 
be the full-scale value of the indicator.   
 

Procedure:

 

Screens shown in this manual represent the displays shown with a 500 
psi indicator.  The IPI will prompt the technician for the appropriate 
pressure at each calibration point. 
 
 
 
 
  
  
  

PSI 

500.00

 

CAL MODE 

Содержание IPI Mk.II

Страница 1: ...Adjustment Manual Industrial Pressure Calibrator JOFRA IPI IPI Mk II...

Страница 2: ...om the factory with all setting access unlocked or available to be changed 2 In supervisory mode each of the parameters can be locked or unlocked using the and keys Select LOC lock for those parameter...

Страница 3: ...y setting a factor of 25 a 40 PSI pressure would display as 1000 40 x 25 The engineering unit displayed on the IPI will be Cust IPI Calibration Procedure Overview Calibration adjustment of the IPI is...

Страница 4: ...needed to connect to the pressure standard Always make sure the hose tubing and fittings etc have a rated working pressure at or above the pressure of the unit Also it is important that there be no l...

Страница 5: ...put 437 50 psi 87 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below...

Страница 6: ...i 50 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressu...

Страница 7: ...psi 12 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the P...

Страница 8: ...PC such as Hyperterminal or PcPlus Connect the RS232 cable to stereo jack on the back of the IPI The other end of the cable should be connected to the terminal PC serial port An adapter may be needed...

Страница 9: ...enu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Second point above ambient Temp Comp 7 First point below ambi...

Страница 10: ...using the serial port use the VAL command Calibration Menu 1 Calibrate Temperature 2 Set Sensor Parameters 3 Calibrate Pressure 4 Reset Temp Compensation 5 First point above ambient Temp Comp 6 Secon...

Страница 11: ...94757 34 473785 35 00 6 894757 3447 3785 3500 437 50 0 06894757 30 164562 30 00 6 894757 3016 4562 3000 375 00 0 06894757 25 855339 26 00 6 894757 2585 5339 2600 312 50 0 06894757 21 546116 22 00 6 89...

Страница 12: ...ment tasks from handheld field instruments to laboratory reference level bench top instruments JOFRA JF Marine Instruments A complete range of calibration equipment for temperature pressure and signal...

Отзывы: