A-6 | Dycor CG1100 Oxygen Analyzer
Variables
Variable Name
Hex ID Notes
Analyzer Type
77
“DYcoR cg1100”
serial number
44
software Version
47
Date
5D
“01/04/2000”
Time
63
“14:36:00”
Baud Rate
25
“9600” or “19200”
node Address
27
“00” see section 1
serial Mode
26
0 = two-wire, 1= four-wire Rs-485
Analyzer name
28
40 character maximum
System Status
5F
Bit
0x0001
0x0002
0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
0x0200
0x0800
0x8000
Function
Alarm 1 high
Alarm 1 low
Alarm 2 high
Alarm 2 low
Memory fail
system failure
Calibration required
Furnace temperature failure
Flow failure
Battery failure
Electronics over temperature
o2 concentration is changing
Held Oxygen Percent
52
Value is held at last reading during calibration
oxygen Percent
08
oxygen PPM
69
Pressure
4D
Requires optional sensor
Flow
6E
Requires optional sensor
calibration state
68
Value
0
0x10
0x11
0x12
0x14
0x20
0x21
0x22
0x24
State
normal operation
Ready to calibrate
calibrating: span valve open. waiting to stabi-
lize
calibrating: Zero valve open . waiting to stabi-
lize
calibrating: Valves closed. waiting to stabilize
Ready to verify
Verifying: span valve open. waiting to stabilize
Verifying: Zero valve open. waiting to stabilize
Verifying: Valves closed. waiting to stabilize
cell Temperature
0B
cell Millivolts
0c
Thermocouple mv
0D
Furnace thermocouple
cPu Temperature
4B
Flow Meter Voltage
4c