Controller / User Interface | 4-1
CONTROLLER / USER INTERFACE
System Start-Up
1.
Plug the power connector from a 24 VDc power supply into the
power entry module on the front of the analyzer.
2.
if you are using an AMETEK 24 VDc power supply (Pn:25469JE), plug
the cord of the power supply into the appropriate Ac power outlet.
NOTE
once you have plugged into the power source, the analyzer will be
ON
. Manually disconnecting the power source will turn the analyzer
OFF
. There is no manual on/off switch or button.
3.
Run either the cal gas or sample gas through the instrument.
4.
Allow the analyzer to warm up for one hour.
5.
set up your alarms, analog and digital outputs, and Pc and device
communications using CG1100 system configurator software. See
instructions in this chapter.
6.
if you are running a cal gas initially, check the o
2
reading by clicking
on the software status tab. if the reading is correct, proceed with run-
ning the sample gas.
7.
if there is an error in the o
2
reading, perform a calibration operation
(chapter 5).
8.
if you are running a sample gas initially, read the sample gas.
Setting up System Parameters
Use the configurator software’s graphical user interface to set up your
analyzer(s). Though you can use the configurator software to set up mul-
tiple analyzers, you can only view the status of a single analyzer at a time.
If using the System 2000 software, refer to the System 2000 manual.
PC Requirements for Configurator Software
• Microsoft Windows 98, NT 4.0 2000 or XP
• 640 x 480, 256-color display