Calibration | 5-1
CALIBRATING THE CG1100-RTP
All oxygen alarms are disabled during calibrations.
Calibration
NOTE
The
span
gas is the high o
2
concentration calibration gas. The
zero
gas is the low o
2
concentration calibration gas.
1.
set up your analyzer for a calibration gas procedure.
see chapter 3 -
Installation
.
2.
using the
Setup
tab of the CG1100 configurator software, enter the
calibration gas values to match the span and zero calibration gases
you intend to use to calibrate your system.
3.
set the 3-way valve on the back of the cg1100 analyzer to
Cal
.
4.
set the 3-way valve that delivers the calibration gases to
Span Gas
.
5.
click the appropriate gas calibration button on the
Calibrate
tab
(
Span & Zero, Span Only, Zero Only
) for dual or single gas calibration.
once the o
2
reading has stabilized on the
Status
display, click the
Calibrate
tab. in the
Calibration
dialog box, click
Calibrate Span
or
Calibrate Zero
to calibrate.
6.
when calibration ends set the 3-way valve that delivers the calibra-
tion gases to the
Off
.
7.
set the 3-way valve on the back of the cg1100 to
Gas
In
.
Aborting a Calibration
Cancel a calibration at any time by clicking the button that you used to
select the type of calibration (
Span & Zero
,
Span Only
,
Zero Only
), or by
clicking
Cancel
at the bottom of the screen.