Power
35
nVision Operation Manual
Power
BATTERY POWER
Power Icon States
The nVision Reference screen display has the following power icon states:
Power Icon Key - nVision Reference Recorder
Icon
Description
External
Power
(USB)
100%
75%
50%
25%
0%
*
Power Remaining
*
Replace Batteries or connect to USB Power
The
icon will appear when the batteries are exhausted and will need to be changed to ensure full functionality of the nVision. Continued use will fur-
ther drain the batteries to a non-operational state where the message “Replace Batteries” will appear across the display. From this state, the only operational
parameter will be the power button . After “Replace Batteries” appears, no measurements will be possible until the batteries are replaced, however, the
recorded data will be preserved.
!
CAUTION:
Never remove battery or USB power when Recording.
!
WARNING:
Do not remove or change the batteries in a hazardous atmosphere.
Automatic Shutoff Timer and Low Power Mode
The Automatic Shutoff Timer is set in CrystalControl. During normal (non-recording) operation, the nVision will power
down when the Automatic Shutoff Timer runs out. The Automatic Shutoff timer will not shut off your nVision while
recording. Instead, The Low Power Mode screen will appear after the Automatic Shutoff Timer runs out. To return to
normal operation simply press the
(
select
)
button.