A-12 | 3050-TE Moisture Analyzer
Table A-1 Continued
Register #0 holds the moisture concentration value not held during verifi-
cation. This value is in ppm.
A pair of registers #135 and #136 provide the same moisture concentra-
tion value in floating point Modicon standard.
Register #1 holds the moisture concentration value held during verifi-
cation cycle. The units of measurement are changing depending of flag
status located in register #65.
A pair of registers #137 and #138 provide the same information in float-
ing point format.
Register #11 is a DataState register, which is designed to synchronise the
data acquisition process. This flag is cleared by read. A value of one indi-
cates new data is available.
Register #8 is the SystemState variable, which is set to alarms and warn-
ings. This value is decoded according to the table below.
Example: The value of SystemState is 12612 which is the same as 3144
hexadecimal number. Corresponding alarms started from the list signifi-
cant bit are: Invalid Reading, Calibration Failure, Flow Alarm, Moisture
Generator Date and Dryer Alarm.
Register #25 is a ModbusCommand register. This register allows sending
special commands to the 3050 analyzer as shown in the table below.
REG
DEFINITIO N
UNITS
SCALING ACCESS
238
ZeroW eekDays (ZeroW eekDays = 84)
-
1
r,w
239
ZeroType (ZeroType = 85)
-
1
r,w
240
M axZeroError (M axZeroError = 86)
hz
100
r,w
241
M axZeroDrift (M axZeroDrift = 87)
hz
100
r,w
242
M oistureSpanSlow (M oistureSpanSlow = 88)
-
1000
r,w
243
FreqO ffsetSlow (FreqO ffsetSlow = 89)
-
1000
r,w
Содержание AMEVision 3050-TE
Страница 9: ...Contents ix...
Страница 16: ...1 6 3050 TE Moisture Analyzer This page left intentionally blank...
Страница 20: ...2 4 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 5 Figure 3 3a Division 1 Installation Dimensions...
Страница 26: ...3 6 3050 TE Moisture Analyzer Figure 3 3b Zone 1 Installation Dimensions...
Страница 27: ...Installation Start Up 3 7 Figure 3 5 Heated Pressure Reduction Flow Figure 3 4 Typical Probe Installation...
Страница 33: ...Installation Start Up 3 13 Figure 3 10 RS 485 Cables Multiple 3050 s...
Страница 34: ...3 14 3050 TE Moisture Analyzer Figure 3 11 RS 485 to RS 232 Coversion for Host PC...
Страница 51: ...PC Interface 4 15 Figure 4 11b Flow Adjust Screen...
Страница 52: ...4 16 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 54: ...5 2 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 60: ...Replacement Parts Service 6 6 This page intentionally left blank...
Страница 84: ...This page intentionally left blank...