3-8   |   3050-TE Moisture Analyzer       

3. XXX DIMENSIONS ARE mm, [XXX] DIMENSIONS ARE INCHES.

2. PRESSURE SET TO 331-358KPa (48-52psig).

1. SET HEATED PRESSURE REDUCER TO 60˚C.

NOTES:

4

7

5

6

3

1

2

LOCATION

VENT TO SAF

E

1/4" TUBE

RELIEF

1/4" TUBE

SAMPLE OUT

(3600psig) max.

24.8MPa

1/4" TUBE

SAMPLE IN

1/2" NPT(M)

REDUCER

PRESSURE

TO HEATED

POWER IN

1/4" TUBE

VAPOR BYPASS

RELIEF

SAMPLE OUT

VAPOR BYPASS

SAMPLE IN

110VAC, 60Hz, 40W max.

 (14AWG)²

3 WIRE, 3.5mm

HEATED PRESSURE REDUCER POWER IN

WHT

BLK

HEATER

{

{

SENSOR

34
21

 [14.25]

362

 [12.25]

311

 [10.00]

254

 [14.75]

375

Figure 3-6  

Heated Pressure 

Reduction Installation.

Содержание AMEVision 3050-TE

Страница 1: ...Model 3050 TE Moisture Analyzer User Manual Process Instruments 455 Corporate Boulevard Newark DE 19702 PN 305670901 Rev Y Configurator Version 2 2...

Страница 2: ...NG USA Delaware 455 Corporate Blvd Newark DE 19702 Tel 302 456 4400 Fax 302 456 4444 USA Oklahoma 2001 N Indianwood Ave Broken Arrow OK 74012 Tel 918 250 7200 Fax 918 459 0165 USA Pennsylvania 150 Fre...

Страница 3: ...al Timing 1 4 Chapter 2 Specifications Operating Environment 2 1 Electrical Classifications 2 1 Ranges 2 1 Outputs 2 1 Data 2 1 Alarms Alerts 2 2 Lower Detectable Limit 2 2 Sensitivity 2 2 Accuracy 2...

Страница 4: ...Electrical Connections 3 8 Analyzer Start up 3 14 Dry Down Period 3 14 Status LEDs and Alarms 3 14 Chapter 4 Parameter Setup Setting Up System Parameters 4 1 Chapter 5 Remote Verification Chapter 6 R...

Страница 5: ...fire electrical shock and injury originally provided by this equipment In addition failure to follow the installation and start up instructions may void the instrument warranty Electrical Safety Up to...

Страница 6: ...the environment and to conserve our natural resources AMETEK recommends that you arrange to recycle this product when it reaches its end of life Waste Electrical and Electronic Equipment WEEE should n...

Страница 7: ...hnique allowing high frequency shield bonding while avoiding the AC ground metal connection In the case of shielded cables the drain wire or braid connection must be kept short A two inch connection d...

Страница 8: ...removing cover turn off all air and sample lines to isolate analyzer from process and wait 5 minutes for pressures to bleed down Difficulty turning cover could indicate pressure build up inside enclo...

Страница 9: ...Contents ix...

Страница 10: ...r on the documentation accompanying the equipment being returned declaring that the goods being returned for repair are American goods the name of the firm who purchased the goods and the shipment dat...

Страница 11: ...gases refer to Table 1 1 The analyzer is calibrated to measure moisture contents from 0 to 2500 ppmv Data output can be in units of ppmv ppmw lb mmscf and mg Nm3 The heart of the 3050 is a quartz cry...

Страница 12: ...by passing a portion of the sample gas through a dryer to remove any moisture present i e the moisture content of the dry reference is less than 0 010 ppmv The difference in the resonant frequency of...

Страница 13: ...ports Verification The 3050 has a built in moisture generator for on board verification A portion of the dry reference gas flows through the moisture generator where a known amount of moisture is adde...

Страница 14: ...nalyzer detects abnormal degradation of sensor proformance over time in the normal mode the analyzer will automatically switch to cell savor mode Once the analyzer switches to cell savor mode it will...

Страница 15: ...nd gaskets used in this instrument oxygen carbon dioxide hydrogen nitrogen helium neon argon krypton xenon methane ethane propane butane ethene ethylene natural gas Freon R12 Freon R22 Freon R114 Carb...

Страница 16: ...1 6 3050 TE Moisture Analyzer This page left intentionally blank...

Страница 17: ...eneral Safety Requirements UL CSA Class I Division 1 Groups BCD T6 Electromagnetic Compatibility Directive EN61326 1 Industrial Low Voltage Directive EN61010 1 Pressure Equipment Directive Russian Gos...

Страница 18: ...ter Accuracy 10 ppb or 10 of reading which ever is greater Moisture Generator Value 1 ppmv nominal Instrument Air Requirements 10 C to 50 C 14 F to 122 F 40 C Dewpoint 40 F Cabinet Cooler 5 5 to 10 3...

Страница 19: ...0 C to 100 C 32 F to 212 F Optimal results are obtained when inlet gas temperature is maintained at 60 C 140 F Power Requirements 120 10 Vac 50 60 Hz 150W max 305 679 901 230 10 Vac 50 60 Hz 150W max...

Страница 20: ...2 4 3050 TE Moisture Analyzer This page intentionally left blank...

Страница 21: ...rding to the customer order and is fused and set for the voltage of the required mains power The power requirements are stated on the metal plate on the side of the casting and in the specification se...

Страница 22: ...pressure reducer recommended at the sample tap Refer to figure 3 4 2 Mount system in selected location and bolt in place Refer to figure 3 3 3 Connect instrument air to 1 4 inch compression fitting Ma...

Страница 23: ...Installation Start Up 3 3 Figure 3 1 TE Sample System Flow Diagram NC NC NO REGULATOR Optional BACK PRESSURE SAMPLE OUT ZERO MODULE DRYER TRACED HEAT INLET SAMPLE DRYER 3050 Dryer Valve...

Страница 24: ...isture Analyzer Figure 3 2 Sample System Components Pilot Valve Sample Out Sample Out 3050 Dryer Zero Module Dryer Back Pressure Regulator Optional Sample In Sample In Instrument Air In Zero Module 30...

Страница 25: ...Installation Start Up 3 5 Figure 3 3a Division 1 Installation Dimensions...

Страница 26: ...3 6 3050 TE Moisture Analyzer Figure 3 3b Zone 1 Installation Dimensions...

Страница 27: ...Installation Start Up 3 7 Figure 3 5 Heated Pressure Reduction Flow Figure 3 4 Typical Probe Installation...

Страница 28: ...NT TO SAFE 1 4 TUBE RELIEF 1 4 TUBE SAMPLE OUT 3600psig max 24 8MPa 1 4 TUBE SAMPLE IN 1 2 NPT M REDUCER PRESSURE TO HEATED POWER IN 1 4 TUBE VAPOR BYPASS RELIEF SAMPLE OUT VAPOR BYPASS SAMPLE IN 110V...

Страница 29: ...utput and alarm contacts from the terminal block to user recording equipment as shown on wiring diagram figure 3 7 2 Connect serial communication from analyzer to the PC being used for cus tomer param...

Страница 30: ...12 TB4 24 13 11 TB3 GND 17 ZONE 1 JUNCTION BOX GND TB4 TX OUT TX IN RX OUT TX IN TX OUT 28 25 28 29 29 PE TB3 28 15 TB3 29 16 28 25 28 29 29 PE TB4 29 16 TB4 25 14 TB4 28 15 TB4 GND 17 CASTING 24 COM...

Страница 31: ...d to TB3 GND If this is not possible for example in case of a ground loop tie the shield at the PC or DCS to chassis and remaining shield to earth through a 0 1uF 500V capacitor 100 to500W 100 to 500W...

Страница 32: ...3 12 3050 TE Moisture Analyzer Figure 3 9 RS 232 Wiring RX TX DTR GND CTS PC type DB9F 3050 OLV type DB9M maximum cable length 10m RX TX DTR GND CTS PC type DB25F 3050 OLV type DB9M...

Страница 33: ...Installation Start Up 3 13 Figure 3 10 RS 485 Cables Multiple 3050 s...

Страница 34: ...3 14 3050 TE Moisture Analyzer Figure 3 11 RS 485 to RS 232 Coversion for Host PC...

Страница 35: ...centration data valid and system alarms In the event of a concentration alarm the red LED will be on The yellow LED reflects sample flow status On indicates sample gas is being measured off indicates...

Страница 36: ...ll AMETEK Service if prob lem persists External temperature should be 80 C or less Call AM ETEK Service Replace moisture generator Replace dryer Review alarm settings Data valid contact opens on all a...

Страница 37: ...iately If it does not start automatically click RUN from the Windows Start menu Type the appropriate drive letter followed by a colon and a backslash and the word setup exe d setup exe and click OK to...

Страница 38: ...e computer serial port settings PC Communications Click the Setup button to configure PC Communications The Serial Port Com munication screen opens Figure 4 2 Communication Protocol Select AMETEK Seri...

Страница 39: ...you have made click Cancel Device Refer to Figure 4 1 Name Type in a name for the analyzer Description Type in a description for the analyzer Save Configuration button Save the analyzer s internal par...

Страница 40: ...erright handcorner Theanalyzer serial number and sensor serial number and the moisture generator and dryer codes are also displayed In the lower right hand corner above the Help button is the Configur...

Страница 41: ...g power if needed Baud Rate Select the baud rate at which data will be transferred Address Identifies the analyzer s address Type the network address for the analyzer being connected RS 485 Identifies...

Страница 42: ...software program PC Communications Once the device communication settings are changed the PC Communications setup screen will automatically open Select the approciate options that correspond to the c...

Страница 43: ...for Modbus Dewpoint Temp This analyzer cannot be configured to accept External Process Pressure via 4 20 mA input The 4 20 mA inputs are to be used for remote verification only Refer to Chapter 5 Sens...

Страница 44: ...20 mA Output Set up your analog output range 20 mA Enter your high analog output limit 4 mA Enter your low analog output limit Hold during Zero Check this box to hold analyzer output during verificati...

Страница 45: ...adjustment at the end of the zero cycle Not checked The analyzer verifies only Ignore Span Drift check box Checked One time span value change is not limited Not checked One time span value change is...

Страница 46: ...nstalled in the analyzer Click OK to accept You must enter a new dryer code each time you replace the dryer Set Moisture Generator Production Code Click to enter the Moisture Generator Production Code...

Страница 47: ...s a 24 hour clock Example 1 00 PM 13 hours Weekly Select the day of the week from the drop down list under Day of the week at the bottom of the box You must also enter a time of day 1 through 24 in th...

Страница 48: ...e As dialog and press Save button The file format is csv which is Excel compatible All data displayed on the Monitor page will be stored in the specified file Off Data collection will terminate and th...

Страница 49: ...er 2 Connect the flow meter to the exhaust line of the analyzer 3 Turn on the analyzer and wait for the analyzer to control the oven tempera ture at around 60 degrees Celsius 4 Then click on the Flow...

Страница 50: ...return to the Monitor tab Test Alarms The Test Alarms push buttons allow you to toggle the alarm contacts to an opened or closed state Use a multi meter set to ohms to read resistance Refer to figure...

Страница 51: ...PC Interface 4 15 Figure 4 11b Flow Adjust Screen...

Страница 52: ...4 16 3050 TE Moisture Analyzer This page intentionally left blank...

Страница 53: ...A current level should be maintained for a minimum of 2 seconds for the analyzer to accept the command Once a command has been accepted the current input must drop below 8 milliamps before a subsequen...

Страница 54: ...5 2 3050 TE Moisture Analyzer This page intentionally left blank...

Страница 55: ...2 Calibrated Sensor Assembly 305 122 903S User Manual 305 670 901 MCU Board 305 110 903S Interface Board 305 113 901S Back Pressure Regulator 305 426 901S Sample System 24 Volt DC Power Supply 230 539...

Страница 56: ...esponsibility of the customer WARRANTY EXTENSIONS AMETEK offers a 2 or 3 year warranty extension for your analyzers The warranty is identical to the original policy supplied with the analyzer SPARE PA...

Страница 57: ...Replacement Parts Service 6 3 Remove hex heps nuts 4 Places Disconnect fittings using a 7 16 wrench Figure 6 2 Tube Fittings Figure 6 1 Hex Keps Nuts Accessing the Analyzer...

Страница 58: ...6 4 3050 TE Moisture Analyzer Disconnect heated sample line connector Figure 6 4 Terminal Block Inside Enclosure Figure 6 3 Sample Line Connector...

Страница 59: ...Replacement Parts Service 6 5 Figure 6 5 Back of Analyzer RS 232 RS 485 In RS 485 Out Out Exhaust 50 PSI Max In Sample Dryer To Dryer From...

Страница 60: ...Replacement Parts Service 6 6 This page intentionally left blank...

Страница 61: ...and the host PC Configuration A set of operating parameters that has been set up using the configurator software for control of a single Model 3050 Mois ture Analyzer Configurator Software A Windows...

Страница 62: ...duces consumption of expensive gases however analyzer response time may be somewhat degraded H2 S Treated Sensor A moisture sensor that has been specially treated for use in sample streams containing...

Страница 63: ...ith an RS 485 interface board installed RS 485 may be run in either 2 wire mode or 4 wire mode over distances up to 1000 meters between the analyzer and the PC or converter Sample Gas that is being me...

Страница 64: ...tions are not possible using two wire mode Units Specific engineering units in which the Model 3050 moisture results are represented Examples are ppmv ppmw etc Verification A check of the accuracy of...

Страница 65: ...The analyzer RS485 connection supports both 4 wire and 2 wire multi drop systems The RS232 connection is used to communicate with a single analyzer using short standard cable The 3050 analyzer underst...

Страница 66: ...RTU protocol mode should be used for monitoring purposes When designing a Modbus RS485 multi drop communication system with the Model 3050 analyzer the system designer should consider the follow ing...

Страница 67: ...d but will not send a respond back to the MODBUS master Communication Parameters The number of data bits is always 8 the baud rate is 9600 or 19200 the number of stop bits and the parity of the analyz...

Страница 68: ...s 0 EXCEPTION CODE The Model 3050 analyzer Modbus protocol implementation supports these exception codes Unsupported function requests from the Modbus master result in excep tion code 01 being returne...

Страница 69: ...ssed on a Model 3050 ana lyzer are scaled and converted into integer values to load into Modbus registers for transmission The register values need to be scaled back at the receiving end to yield the...

Страница 70: ...ssure analog input Analog0 64 Hex counts 1 r 13 Flow analog input Analog1 65 Hex counts 1 r 14 Sensor temperature analog input Analog2 66 Hex counts 1 r 15 Process pressure analog input Analog3 67 Hex...

Страница 71: ...ow filter coefficient FlowFilter 0B Hex 100 r w 41 Sensor temperature filter coefficient 100 r w CellTempFilter 0C Hex 42 Process pressure filter coefficient 100 r w CustomerFilter 0D Hex 43 Current o...

Страница 72: ...ange Alarm2Hi 2E Hex 1 r w 69 Concentration alarm range Alarm2Lo 2F Hex 1 r w 70 Process pressure units PressUnits 30 Hex 1 r w 71 Bypass valve state BypassState 31 Hex 1 r w 72 Zero coefficient Frequ...

Страница 73: ...oefficient MoistureSpan 6E Hex 1000 r w 98 Track or hold current output flag HoldOut 71 Hex 1 r w 99 Fast cycle 0 slow cycle 1 SlowTiming 72 Hex 1 r w 100 Minimum verification duration min 1 r w VerDu...

Страница 74: ...ficient PolyCoeff0 32 Hex float r 125 126 Sensor calibration coefficient PolyCoeff1 33 Hex float r 127 128 Sensor calibration coefficient PolyCoeff2 34 Hex float r 129 130 Sensor calibration coefficie...

Страница 75: ...2 Current time string Time 5E Hex string r w 213 220 Analyzer model name ModelName 77 Hex string r 221 225 Dryer production code DryerDateCode 79 Hex string r w 226 228 Reserved string 229 Hi Span Lim...

Страница 76: ...ead A value of one indi cates new data is available Register 8 is the SystemState variable which is set to alarms and warn ings This value is decoded according to the table below Example The value of...

Страница 77: ...ng PPM Hours Concentration Alarm Bit0 Bit1 11 Concentration out of range Invalid Reading Bit2 11 Verification or System Alarm Process pressure Bit3 11 Process pressure out of limits VALUE CO M M AND D...

Страница 78: ...r Model 3050 analyzer 1 byte Run Status FFh for Analyzer on Line invalid signal 0 00h for Analyzer off Line invalid signal 1 2 bytes Status Word System State which is register 8 The most significant b...

Страница 79: ...section All of the configuration parameters of the analyser can be modified by the write one or multiple holding registers command In most cases it is a one step operation involving setting the conten...

Страница 80: ...er address 66 holding register 40067 Value 1000 Example 3 Setting the Low Alarm Limit TASK Set the analyser to produce a low concentration alarm when the mois ture concentration falls below 0 ppm ACTI...

Страница 81: ...06 Register address 46 holding register 40047 Value 100 multiplied by scale of 100 Example 7 Switching to Sensor Saver Mode TASK Set the analyser to operate in the sensor saver mode ACTION Write one...

Страница 82: ...e multiple holding registers function 16 Register address 162 holding register 40163 value 4169Hex Register address 163 hold ing register 40164 value 7200Hex Air0 ACTION 2 Set gas related coefficients...

Страница 83: ...ill be set to perform a verification cycle on the 3rd day of each month at noon 12 00 ACTION 1 Set the verification type to monthly Write one holding register function 06 Register address 81 holding r...

Страница 84: ...This page intentionally left blank...