4-14 | 3050-TE Moisture Analyzer
Enter the value displayed on the external flow meter and click the
Update button
.
The
External
field will lock again and the configurator will calculate a new Flow
Span and send it to the analyzer. Then the analyzer will attempt to control the flow
again and when it’s done, the
External
field will unlock again. You can repeat this
process until the value of the
Internal
flow meter matches the external flow meter.
When calibration is complete press OK, to return to the
Monitor
tab.
Test Alarms
The Test Alarms push buttons allow you to toggle the alarm contacts to an opened
or closed state. Use a multi-meter set to ohms to read resistance. Refer to figure 2.7
for location of contacts on analyzer.
· Opened contacts should read infinity.
· Closed contacts should read zero.
Test mA Output
The Test mA Output push buttons allow you to test the analog outputs. Use a meter
to test output. Refer to figure 2.8 of for wiring.
To discontinue test mode, press another tab. Test mode will automatically time out
after 10 minutes of inactivity.
When the analyzer switches from test mode to online, the analyzer
resets itself.
Figure 4-11a.
Flow Adjust Screen.
NOTE
Содержание AMEVision 3050-TE
Страница 9: ...Contents ix...
Страница 16: ...1 6 3050 TE Moisture Analyzer This page left intentionally blank...
Страница 20: ...2 4 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 5 Figure 3 3a Division 1 Installation Dimensions...
Страница 26: ...3 6 3050 TE Moisture Analyzer Figure 3 3b Zone 1 Installation Dimensions...
Страница 27: ...Installation Start Up 3 7 Figure 3 5 Heated Pressure Reduction Flow Figure 3 4 Typical Probe Installation...
Страница 33: ...Installation Start Up 3 13 Figure 3 10 RS 485 Cables Multiple 3050 s...
Страница 34: ...3 14 3050 TE Moisture Analyzer Figure 3 11 RS 485 to RS 232 Coversion for Host PC...
Страница 51: ...PC Interface 4 15 Figure 4 11b Flow Adjust Screen...
Страница 52: ...4 16 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 54: ...5 2 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 60: ...Replacement Parts Service 6 6 This page intentionally left blank...
Страница 84: ...This page intentionally left blank...