3-16 | 3050-TE Moisture Analyzer
Zero parameter is out of
range
Sample sensor hardware
failure
Replace zero dryer or call
AMETEK Service.
Replace sensor or call AM-
ETEK Service.
Call AMETEK Service.
* Indicates System Alarm and Data Invalid Signal
Oven temperature is out of
tolerance.
This will occur during start-
up until the oven warms
up. Call AMETEK Service if
problem persists.
Check inlet and outlet pres-
sure. Call AMETEK Service if
problem persists.
Call AMETEK Service.
Sample flow rate too high or
too low
Battery needs to be replaced.
Analyzer detected problem
with reference gas.Excessive
internal temperature.
Moisture generator date has
expired.
Check and/or replace dryer.
Call AMETEK Service if prob-
lem persists
External temperature should
be 80
°
C or less. Call AM-
ETEK Service.
Replace moisture generator.
Replace dryer
Review alarm settings.
Data valid contact opens on all alarms and stays closed during normal
functions and readings. An open data valid contact indicates verifica-
tion is in process or an alarm condition. Red LED flashes 11 times if
an invalid condition occurs.
Table 3.1 LEDs and Alarms
Zero Alarm Problem*
1
Sample Sensor Failure*
2
Flow Out of Tolerance*
5
Battery Low*
6
Reference Dryer Alarm* 7
Enclosure Temperature
8
Moisture Generator Date N/A
Alarm Source
LED Flashes per Cycle
Problem
Action
Analyzer performance out of
tolerance as detected during
verification cycle.
NOTE
Dryer Alarm
10
Dryer failure Imminent
Moisture concentration is out
of user defined limits.
Concentration Alarm
11
Calibration Failure*
3
Oven Temperature*
4
Содержание AMEVision 3050-TE
Страница 9: ...Contents ix...
Страница 16: ...1 6 3050 TE Moisture Analyzer This page left intentionally blank...
Страница 20: ...2 4 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 5 Figure 3 3a Division 1 Installation Dimensions...
Страница 26: ...3 6 3050 TE Moisture Analyzer Figure 3 3b Zone 1 Installation Dimensions...
Страница 27: ...Installation Start Up 3 7 Figure 3 5 Heated Pressure Reduction Flow Figure 3 4 Typical Probe Installation...
Страница 33: ...Installation Start Up 3 13 Figure 3 10 RS 485 Cables Multiple 3050 s...
Страница 34: ...3 14 3050 TE Moisture Analyzer Figure 3 11 RS 485 to RS 232 Coversion for Host PC...
Страница 51: ...PC Interface 4 15 Figure 4 11b Flow Adjust Screen...
Страница 52: ...4 16 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 54: ...5 2 3050 TE Moisture Analyzer This page intentionally left blank...
Страница 60: ...Replacement Parts Service 6 6 This page intentionally left blank...
Страница 84: ...This page intentionally left blank...