Schedule Tab
Use the
Schedule
tab to schedule routine
Zero
.
NOTE
Figure A-5.
Schedule tab.
Abort
Button
Click to terminate verification or zero.
Zero Now Button
Click to begin zero cycle now.
Adjust Offset check box
Checked
The analyzer performs an offset adjustment at the end
of the zero cycle.
Not checked
The analyzer verifies only.
Set Clock
(Time Synchronization) Click
Set Clock
to synchronize the clock within the
analyzer with the PC. The
Time Synchronization
box opens with the PC
time and date and the analyzer time and date. Click S
ynchronize
to set the
time, or click
Cancel
to close the box.
Set Dryer Production Code
Click to enter the
Dryer Production Code
for the dryer that is installed in
the analyzer. Click
OK
to accept.
You must enter a new dryer code each time you replace the dryer.