Installation & Start-Up | 3-11
Electrical Connections (Figure 3-6)
This analyzer has an internal AC voltage selector that must be set
correctly to prevent damage. Only qualified service personnel may ac-
cess the interior of the analyzer. If qualified service personnel are not
available, contact AMETEK service.
Voltage selection
The 3050-AMS analyzer comes from the factory set to operate on 120 VAC.
If you must switch the setting to 240 VAC (or back to the 120 VAC setting),
follow the procedure below.
You must follow this procedure to prevent damage to/malfunction of
the analyzer.
1. Ensure that the area is non-hazardous and the analyzer has been de-
energized and power disconnected
2. Remove the terminal cover on back panel.
3. Select 115 or 230 VAC on the voltage selector (10) located on the top
left-hand side of the user interface board.
4. Replace terminal cover.
5. Clearly mark the operating voltage on the rear panel analyzer serial
number label in the lower right-hand corner.
ON
ON
TB3
TB2
TB1
ANALOG
OUTPUT
DATA
VALID
CONC
ALARM
SYSTEM
ALARM
OUT
- +
IN
RD TD
+ - + -
LINE VOLTAGE
FUSE
L N G
1 2 3
RS-485
TERMINATOR
1 2 3 4 5 6 7 8 9 10 11 12 13 14
115
230
Voltage
Selection
Switch
10
Figure 3-6.
115 / 230 voltage selection switch.
!
WARNING
!
CAUTION
Содержание 3050-AMS
Страница 12: ...xii 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 22: ...2 4 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 3 Figures 3 1c Model 3050 AMS dimensions top and sides...
Страница 42: ...3 20 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 64: ...5 10 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 66: ...6 2 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 84: ...7 18 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 96: ......