Installation & Start-Up | 3-19
Procedure for speeding up the dry down period
To decrease the time required to dry down the 3050-AMS analyzer, follow
this procedure:
1. Make all gas connections and apply power to the analyzer. Allow the
analyzer to warm up and run with the factory-default settings for a
period of 6 to 12 hours.
2. At the end of this interval, set as follows:
Zero Schedule Type
to “Daily”
Zero Duration
to 720 cycles (i.e., 12 hours)
Zero Hour
to 1
The exact menu location for each of these parameters can be found by
referring to the Menu Map at the beginning of Chapter 4.
Once you apply these settings, the analyzer will run a 12-hour zero
every day at 1 AM. The analyzer will spend 50% of the time on the
sample gas and 50% of the time on the internal zero. These settings
enable the analyzer to more effectively dry all of the surfaces within
the sample manifold.
3. After 3 to 4 days, the analyzer setup can be returned to normal opera-
tion (e.g., the factory default values).
Performing an Initial Zero
To start the initial zero cycle:
1. Press the
Zero
key.
2. Press the #8 key (arrow down) until you see “Zero” between the ar-
rows. Select “Zero” by pressing
Enter
.
This initial zero cycle takes approximately two hours. When the zero cycle
is complete, the analyzer will adjust the offset and automatically return to
normal sampling.
The 3050-AMS analyzer requires time to fully recover to normal
operation after zeroing. During the recovery period, “Equilibrating”
will be shown on line two of the display.
NOTE
Содержание 3050-AMS
Страница 12: ...xii 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 22: ...2 4 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 25: ...Installation Start Up 3 3 Figures 3 1c Model 3050 AMS dimensions top and sides...
Страница 42: ...3 20 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 64: ...5 10 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 66: ...6 2 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 84: ...7 18 3050 AMS Moisture Analyzer This page intentionally left blank...
Страница 96: ......