2 series specifications
LSP
HD
20
LSP
HD
21
LSP
HD
22
Operating wavelength
2.2 µm
1.9 µm
1.9 µm
Temperature range
200 to 850 °C
392 to 1562 °F
300 to 1000 °C
572 to 1832 °F
400 to 1200 °C
752 to 2192 °F
System Accuracy
(1)
(5 to 95% of range)
± 2 °C / 3.6 °F
Focus distance
(2)
1m / 39.7in to infinity (continuously adjustable)
Field of View
(static to 95% of radiance)
300:1
(0.0033
radians)
500:1
(0.002 radians)
500:1
(0.002 radians)
Repeatability
< 0.5°C / 0.9°F
Temperature Resolution
(3)
< 5% of range
(noise)
< 15% of range
< 30% of range
Above 50% of range
≤ 6°C / 10.8°F
≤ 6°C / 10.8°F
≤ 2°C / 3.6°F
≤ 1°C / 1.8°F
≤ 6°C / 10.8°F
≤ 6°C / 10.8°F
≤ 4°C / 7.2°F
≤ 1°C / 1.8°F
≤ 6°C / 10.8°F
≤ 6°C / 10.8°F
≤ 4°C / 7.2°F
≤ 1°C / 1.8°F
Scan Angle
(4)
80°
Scan Rate
Factory set
User adjustable
50 Hz
10 to 150 Hz in 1 Hz steps
Speed of response
(Tau - linear output on rising edge)
≤ 1.5µs
≤ 1µs
≤ 1µs
Ambient Stability
deltaT[indicated] / deltaT[ambient]
≤ 1°/10°
(1) Includes error correction
(2) Focus is factory set to 2m
(3) 4 line averaging below 15% of range
(4) Scan angle may be set in the range 40 to 80° by the PC software