![Ametek Land NIR-2K Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/ametek-land/nir-2k/nir-2k_installation-manual_2931436015.webp)
Installation Guide
Thermal Imaging Camera
NIR-656 / NIR-2K
11
The TRACO Power Supply Unit is supplied with
a 3-pin AC input connector (Fig. 4-1). You will
need to disassemble this connector to install the
cable.
1) Unscrew the cover from the mains
Input
on the Power Supply Unit.
Fig. 4-1
Fig. 4-2
Fig. 4-3
2) The easiest way to disassemble the 3-pin
AC input connector is to first screw the
connector assembly onto the input socket
(Fig. 4-2).
3) Grip the rear section of the input connector
and unscrew it from the Power Supply Unit
(Fig. 4-3).
Fig. 4-4
The connection terminals inside the input
connector will now be revealed (Fig. 4-4).
4) Unscrew the connector front section from
the Power Supply Unit and disassemble
the remaining parts of the connector.
5) Slide the rear locking ring, gasket and
shell body over the mains input cable.
4.1 Connecting the mains input to the TRACO Power Supply Unit