Preparing the DALT Tank for Electrophoresis
Hoefer DALT System
26
Amersham Biosciences
Figure 14.
Raising the
barrier combs to allow
buffer circulation
5. Make a note on a label on the tank lid when you change the buffer.
Although it is possible to use each tank of buffer for 2-3 runs, by mixing the center
and cathode [left] chambers between runs, it is better to change the buffer before
every run. Fresh buffer ensures more consistent results.
When a tank is not going to be used for several days, siphon out the buffer, rinse the
tank with water, and allow the tank to sit empty. This prevents growth of bacteria
and molds which can occur in used buffer. Such contamination can give rise to
“mysterious” protein bands across a whole set of gels.
6. When the buffer is distributed in all chambers, replace the barrier combs in the
outer slots to accommodate the 20
×
25 cm cassettes.
Figure 15.
Top view of
DALT Tank, with barrier
combs in place
When properly positioned, the ends of the silicone rubber flaps face the outer walls
of the tank (Figure 15).
Pump
Cassette
Outer barrier comb slot,
accommodates 20
×
25 cm