Performing a control Solution test
1. insert the test strip to turn on the
meter
Insert the test strip into the meter. Wait
for the meter to display the test strip
and blood drop symbols.
2. Press M button to mark this test as a control solution
test
With “QC” displayed, the meter will not store your test
result in memory. If you press M button again, the “QC”
will disappear and this test is no longer a control solu-
tion test.
WArninG:
When doing the control solution test, you have to mark it so that the
test result will not be stored in the memory. Failure to do so will mix up
the blood glucose test results with the control solution test results in
memory.
3. Apply control Solution
Shake the control solution vial thoroughly before
use.
Squeeze out a drop and wipe it off, then squeeze out
another drop and place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of the test strip to touch
the drop. Once the confirmation window fills completely, the meter will
begin counting down.
To avoid contaminating the control
solution, do not directly apply
control solution onto a strip.
4. read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and
it should fall within this range. If not,
please read the instructions again and
repeat the control solution test.
Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do
NOT test your blood. Contact the local customer service or place of
purchase for help.
note:
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood
glucose level.
• See the Meter Maintenance section for important information about
your control solutions.