8.0 Technical data  

 

 

FMT-310A5 

FMT-310B5 

FMT-310C5 

Max. forc 

Pressure and tension 

5 N 

50 N 

500 N 

Stroke 

Maximum 

400 mm 

Speed 

 

0,1...900mm/min 

Accuracy 

Force 

+/- 0,5% MW (+/- 1 digit rounding error) 

Path 

+/- 0,01 mm (+/- 1 digit rounding error) 

Speed 

+/- 0,5% (+/- 1 digit rounding error) 

Resolution 

Force 

0,001 N 

0,01 N 

0,1 N 

Path 

0,01 mm 

Speed 

0,1 mm/min 

Measurment rate 

Intern 1000Hz | Communication 100Hz 

Working chamber 

Depth 

120 mm (Center tool fitting until  Antriebssäule) 

Additional height 

0 ... 240mm (adjustable in 40mm steps) 

Objektträgerplatte 

270 mm x 300 mm (0ptional) 

Tool fitting 

Load cell 

M4 

M6 Screw thread 

Objektträgerplatte 

M6 threaded hole 

Control 

Indication /in-put 

5,7“ LCD Monitor mit Touchpanel 

Controller 

PLC | 16bit | 20Mhz 

Interface 

RS232C, Baudrate 19200 

Current supply 

 

85…264 VAC (50...60Hz) | 1,8A @ 230VAC 

Temperatur range 

Operation 

10...40° C 

Storage/Transport 

-10...60° C 

Weight 

 

ca. 20,5 kg (Control unit :ca. 1,5kg) 

Dimensions 

L x W x H 

400 mm x 300 mm x 900 mm 

 

Additional  technical  data  for  the  test  stand  with  optional  equipment  may  be  mentioned  within  the 
enclosure 2. 

9.0 Maintenance 

9.1 Linear bearing 

The  linear  bearing  has  to  lubricated  periodical,  depending  on  the  frequency  of  use  and  the 
environmental conditions but at least after 6 month. Therefore please open the upper bellows on the 
load cell. In front of the drive screw you can see a lubricating nipple. Lubricate exclusively with lithium 
soap grease DIN-Typ KP2-K. Attention! Lubricants with solid additives (z.B. MoS², Graphit oder PTFE) are 
not  appropriate  for  the  use  within  the  test  stands.  You  can  order  an  adequate  lubricating  grease 
(Art.No.:  FMT-9802)  and  a  manual  compactor  with  a  fitting  tube  attachment  (Art.No.:    FMT-9803)  on 
www.alluris.de . 

9.2 Updates (Software) 

Software-Updates can be brought in via the seriell interface of a PC. It is advisable to save the all the 
date before initiating the update as it can happen that parameter sets and result list get canceld during 
the update. Excat instructions about how to effect an update you’ll recieve with the data medium. 

Содержание FMT-310

Страница 1: ...1 Bedienungsanleitung Operation Manual FMT 310 Force Tester BDA Version 1 1 DE ENG ...

Страница 2: ...N 10 5 3 3 PREDEFINED FUNCTIONS 10 5 3 3 1 ZERO POINT SEARCH 11 5 3 3 2 BREAKDETECTION 11 5 3 3 3 FORCE REGULATION 11 5 3 3 4 PATH REGULATION 11 5 3 3 6 HARDNESSGRADE 11 5 3 4 PREDEFINED TEST PROGRAMMS 12 5 3 4 1 BREAK TESTING 1 DIAGRAMM 13 5 3 4 2 BREAK TESTING 2 PEAK VALUES 13 5 3 4 3 BEND TESTING PATH DEPENDING 14 5 3 4 4 ENDURANCE TESTING PRESSURE WITH PRESSURE FORCE REGULATION 15 5 3 4 5 TOW ...

Страница 3: ... only The instrument is equipped with a temperature compensation for 0 40 C Use the instrument in this temperature range only If you effect a material testing during which the material can break or sliver in a way that people the test stand or other instruments in the near could get damaged you have to install protective equipment If you have no protective equipment at your disposal we would be gl...

Страница 4: ...test has been finished succesfully the change to the main menu takes place Press now the button REF it takes place an automatic adjustment of the internal zero reference machine zero For this the load cell gets driven into the highest position After the reference path has been effected succesfully the test stand is ready for use Now you can choose the wanted function within the main menu In depend...

Страница 5: ...agram Break testing 2 Diagram Setting Results Bend testing Path depending Settings Results Diagram Endurance testing Force regulation Settings Results Diagram Tow testing 1 Path depending Settings Results Diagram Tow testing 2 Peak Settings Results Stretch testing 1 Path depending Settings Results Diagram Stretch testing 2 Force regulation Settings Results Diagram Service adjust date time report m...

Страница 6: ...el of menu Scroll To jump line by line within the list of parameters or results ATTENTION within the manual mode the arrows indicate only the movement of the slide per direction List of parameter After pressing the button appearce a list of parameteriseable settings Enter keyboard By pressing the button the input box to change the parameterizeable settings appearse Jump first last line The arrow k...

Страница 7: ...of operation 5 1 Explanation of instrumentation 5 1 1 Force measurement For the force measurement high precision measuring beam get used their deflection during force transmission gets recorded by a extension measuring tape The conversion of the analog signals into digital form takes places within the load cell The digital signals of the D A converter constitutes 1000 Hz the transfer rate to the c...

Страница 8: ...nspection process gets aborted and an infotext display takes place in the lower line of the display The load cells have an additional electronic overload protection which takes effect independently of the setted limits as soon as a damage of the load cell is imminent The control unit tries automatically to take it off into the diametrical direction with a propulsion command Similarly the drive doe...

Страница 9: ...hieved 5 2 4 Ride in the Tipp mode Tipp In the tip mode which gets setted by the mode button the drive gets moved with a preselected speed into the required direction till the button gets released or a limit gets achieved 5 2 5 Indication and reset of the peak values maximum values The absolut maximum values for traction force and compression force get saved within the load cell The two maximum va...

Страница 10: ...ell all parameter sets in all test programms get reset onto the default values to avoid possible damage on the construction If you change the tools in use or the material to check it might be necessary to adjust the parameter to avoid damages on the construction The submenu to regular the parameter gets called by pressing the parameter button 5 3 2 Definition of the valve position Reference positi...

Страница 11: ...ation a predefined force gets kept for kept for a predefined amount of time The drive re enacts automatically the position of the load cell provided that the material property demands this The basic and the driving speed depends of the regulated hardnessgrade of the material Is the hardnessgrade understated it can lead to a consequently overshooting of the regulation Is the hardnessgrade overstate...

Страница 12: ... readings of measurment of force Minimise y axle with readings of measurment of force Enlarge x axle with readings of measurment of length Minimise x axle with readings of measurment of length The recorded force values are placed on the y axle the corrisponding position values on the x axle The scaling of the diagram takes automatically place on the base of the size of the values and the nuber of ...

Страница 13: ...ut on 0 Reference Spee vHome 0 1 mm min 900 0 mm min Hardnessgrade measurment object 0 99 START position HOME position 405 00 Speed bread detection 0 1 mm min 900 0 mm min Force Breakdetection 0 00 100 Speed till break 0 1 mm min 900 0 mm min Force break 0 00 100 5 3 4 2 Break testing 2 Peak values The inspection process serves to appraise the breaking force at various successive tests by pressing...

Страница 14: ...orce and path 9 Automatically the slide gets moved upwards with the speed vHome to the HOME position Adjustable parameter for the test procedure Bend testing path dependend Parameter Permissible range of values Reference Min Max Setpoint position DOWN SP top 405 00mm Absolut on 0 Reference Setpoint position TOP 0 00mm 405 00mm Absolut on 0 Reference Setpoint force PRESSURE 0 N Nominal range of loa...

Страница 15: ...er 6 Engaging of the force regulation and movement of the load cell until achieving the holding force Remaining for the predetermined pause time after the first arrival at the holding force 7 End of the reading record and ride to the Home positionsition Adjustable parameter for the test procedure Load force regulated Parameter Permissible range of values Reference Min Max Setpoint position DOWN SP...

Страница 16: ...d VStart 0 1 mm min 900 0 mm min Force Break detection 0 00 100 Speed break 0 1 mm min 900 0 mm min Force break 0 00 100 5 3 4 6 Tow testing 2 Peak values Like the former one the inspection process serves to appraise the tear out force at a continuous tension at a uniform speed until the break of the measurment Instead of a test serie the peak values and position of the measurment object at the mo...

Страница 17: ...or the pause time T 6 Return for the path_2 absolut relating to 0 position test ATTEND prefix of the way value 7 End of the reading record and return to the HOME position Adjustable parameter for the test procedure Strech testing path depending Parameter Permissible range of values Reference Min Max Setpoint position DOWN SP top 405 00mm Absolut on 0 Reference Setpoint position TOP 0 00mm 405 00mm...

Страница 18: ... deposit too The description of the operation may be described within the enclosure 1 Parameter Zulässiger Wertebereich Reference Min Max Setpoint position DOWN SP top 405 00mm Absolut on 0 Reference Setpoint position TOP 0 00mm 405 00mm Absolut on 0 Reference Setpoint force Pressure 0 N Nominal range of load cell Setpoint force TENSION 0 N Nominal range of load cell HOME position SP top SP down A...

Страница 19: ... becomes activated after the next initiation of the instrument 6 1 2 Password Categorical the servicefunctions are protected by a password You can change these passwords and afterwards not even the Alluris Service is able to access the servicefunctions without a direct interference into the firm ware It s possible to protect the parameters for the corrisponding inspection processes with a user pas...

Страница 20: ...rence has been setted the measurment range of the nominal value gets newly setted by adherencing an adapted reference standard Only when the new values have been exactly averaged out and checked on their plausibility within a width which is predetermined within the force gauge the new settings become assumed Otherwise the old values remain stored After every force calibration it is necessary to sw...

Страница 21: ...memory can be necessary especially if the memory capacity is exhausted Within most of the menues the remaining memory capacity gets indicated within the info line Before cancelling the data you can import them into Exel by the help of the Fmt_Connect_Software in this way you can save them everlasting 7 0 Connection to the PC The Force Tester FMT 310 can be conected to the serial interface of a PC ...

Страница 22: ... x H 400 mm x 300 mm x 900 mm Additional technical data for the test stand with optional equipment may be mentioned within the enclosure 2 9 0 Maintenance 9 1 Linear bearing The linear bearing has to lubricated periodical depending on the frequency of use and the environmental conditions but at least after 6 month Therefore please open the upper bellows on the load cell In front of the drive screw...

Страница 23: ...a comunication mistake between the load cell and the control unit Pleas check if the two cable for the load cell 9 p D Sub switch and M8 sensor connector are correctly connected Switch off the test stand to reset all the functions Try again to initiate the test stand If the load cell is defekt it can be necessary to change the whole force measurment unit see in addition chap 7 4 Our service can re...

Страница 24: ... return it to Alluris for recycling For more information please contact our website www alluris de The compliance to the requirements of all relevant EEC directives is confirmed by the CE marking of the product Calibration Confirmation We hereby confirm in accordance to DIN EN 10204 2 1 that this instrument has been tested in accordance with DIN EN 9000 2000 approved procedures The instrument meet...

Страница 25: ...25 Service Adressen Alluris GmbH Co KG Technischer Service Basler Strasse 63 DE 79100 Freiburg Deutschland Fon 49 0 761 47979 3 Fax 49 0 761 47979 44 service alluris de www alluris de ...

Отзывы: