Instructions for Use for KardiaMobile Card (AC-021) 

 

 

 

 

 

 

 

 

 

 

 

Page 14 of 17 

 

Electrical Safety 

Guidance and manufacturer's declaration - electromagnetic emissions 

KardiaMobile Card is intended for use in the electromagnetic environment specified below. The customer or the 

user of KardiaMobile Card should assure that it is used in such an environment. 

Emissions test 

Compliance  Electromagnetic environment - guidance 

RF emissions 

CISPR 11 

 

Group 1 

 

KardiaMobile Card uses RF energy only for its internal 

function. RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment. 

 

RF emissions 

CISPR 11 

Class B 

KardiaMobile Card is intended for use in domestic 

surroundings. 

Harmonic emissions 

IEC 61000-3-2 

N/A 

KardiaMobile Card is powered from a lithium manganese 

dioxide battery and does not require AC mains power 

Voltage fluctuations /  

flicker emissions 

IEC 61000-3-3 

N/A 

 

Guidance and manufacturer’s declaration—electromagnetic immunity 

KardiaMobile Card is intended for use in the electromagnetic environment specified below. The customer or the 

user of KardiaMobile Card should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment - 

guidance 

Electrostatic Discharge 

(ESD) 

IEC 61000-4-2 

±2 kV contact 

±4 kV contact 

±6 kV contact 

±8 kV contact 

±2 kV air 

±4 kV air 

±8 kV air 

±15 kV air 

±2 kV contact 

±4 kV contact 

±6 kV contact 

±8 kV contact 

±2 kV air 

±4 kV air 

±8 kV air 

±15 kV air 

Floors should be wood, concrete, 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%. 

Electrical fast 

transient/burst 

IEC 61000-4-4 

N/A 

N/A 

KardiaMobile Card is powered 

from a lithium manganese dioxide 

battery and does not require AC 

mains power. 

Surge 

IEC 61000-4-5 

N/A 

N/A 

Voltage dips, short 

interruptions, and 

voltage variations on 

power supply input 

lines 

IEC 61000-4-11 

N/A 

N/A 

Power frequency 

(50/60 Hz) 

magnetic field 

IEC 61000-4-8 

30 A/m 

30 A/m 

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment. 

Содержание AC-021

Страница 1: ...trademarks of AliveCor Inc in the United States and other jurisdictions Instructions for Use IFU for KardiaMobile Card AC 021 21LB01 01 December 2021 AliveCor Inc 189 N Bernardo Avenue Suite 100 Moun...

Страница 2: ...ionality 6 Setting up your KardiaMobile Card device for the first time 7 Recording an EKG 7 EKG Analysis 8 Heart Rate 10 Environmental Specifications 10 Expected Service Life 10 Maintenance 10 Device...

Страница 3: ...your heart rhythm is provided upon completion of your EKG recording See the EKG Analysis section for more details 4 KardiaMobile Card requires a compatible smartphone or tablet and the Kardia app a T...

Страница 4: ...ny EKG result including normal You should notify your physician for possible changes in your health or if you are experiencing any concerning symptoms 2 Chest pain or pressure is a medical emergency H...

Страница 5: ...making 12 The device has not been tested for and is not intended for pediatric use 13 Device contains a lithium manganese dioxide battery that is not removable or replaceable 14 Do not use the electr...

Страница 6: ...ard System is intended for use by healthcare professionals patients with known or suspected heart conditions and health conscious individuals The device has not been tested and is not intended for ped...

Страница 7: ...evice list at www alivecor com compatibility 3 Make sure Bluetooth is turned on in your smartphone or tablet settings 4 Launch the Kardia app and tap Create Account 5 Follow the on screen instructions...

Страница 8: ...G recording is complete 8 The device will turn off automatically after use EKG Analysis Upon completion of your EKG recording the EKG is processed by AliveCor s Instant Analysis algorithms in the Kard...

Страница 9: ...recordings shorter than 30 seconds Record a new EKG Try to relax and hold still rest your arms or move to a quiet location that will allow for a full 30 second recording Unclassified Atrial fibrillat...

Страница 10: ...favorably compared to Lead I recordings of a standard FDA cleared 12 lead device Clinical equivalence of recordings from these Kardia devices and the 12 lead EKG device were also validated by board c...

Страница 11: ...ces KardiaMobile Card has been tested and deemed in conformance with the relevant requirements in IEC 60601 1 2 2014 Class B for Electromagnetic Compatibility EMC FCC and Industry Canada IC Compliance...

Страница 12: ...s open the KardiaMobile Card device is selected and is ready to pair Make sure the amber LED light turns on when you press the white power button located inside the A symbol on the right corner of the...

Страница 13: ...recording Ensure that the Enhanced Filter is on Make sure Mains Filter is set appropriately for your geographical location This can be adjusted under the Kardia app Settings 3 On my EKG the recording...

Страница 14: ...immunity KardiaMobile Card is intended for use in the electromagnetic environment specified below The customer or the user of KardiaMobile Card should assure that it is used in such an environment Im...

Страница 15: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symb...

Страница 16: ...ated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1...

Страница 17: ...Symbol Interpretation Manufacturer Refer to instruction manual booklet Read instructions before use Do not dispose with household waste Temperature range Do not cut shred or attempt to destroy device...

Отзывы: