11.3.4 Storage and maintenance
Visually inspect the cradle for any noticeable damage, for example, deterioration such as discoloration, pitting, or
cracks.
Warning:
Do not use the cradle if any damage is detected. Please contact iTero Customer Support for further
instructions.
11.4
Cleaning and disinfecting the scanner touch screen and the wheel-stand handle
The scanner screen and wheel-stand handle must be cleaned between patients, as follows:
1. Clean all outer surfaces using approved disinfectant wipes, or approved disinfectant liquid with a clean lint-free
wipe, and follow the manufacturer’s instructions. For a list of approved materials, see
2. Remove any residual liquid disinfectant with a clean lint-free wipe.
Note:
Do not use abrasive cleaners and/or corrosive cleaning agents or disinfectants with acids, bases, oxidizing
agents, or solvents.
11.5
General cleaning
All scanner parts and accessories not outlined above should be cleaned according to standard operating
procedures or local regulations.
iTero Element 5D and iTero Element 5D Plus imaging systems User manual
© 2022 Align Technology, Inc. All rights reserved.
191