Instrument functions
R&S
®
NGU
52
User Manual 1179.2531.02 ─ 03
● Source: The R&S
NGU goes to source mode, current flows out from the instru-
ment.
Note:
If "Sink" mode is selected, the source current is limited to 1 mA. Otherwise, if
"Source" is selected, the sink current is limited to 1 mA.
3. Select "Set" to configure the trigger condition.
6.3
Current priority mode
Current priority mode is a special mode that the R&S
NGU operates in to protect cur-
rent sensitive DUTs. By default, the R&S
NGU operates in voltage priority mode.
For more information, see
Chapter 5.5.1, "Voltage and current priority modes"
1. Press [Settings] key.
The R&S
NGU displays the device/channel menu window.
2. Select the channel tab to set current priority mode.
The R&S
NGU displays the channel menu.
Figure 6-12: Channel menu
3. Activate the "Current Priority Mode" menu item.
The "CPM" is highlight in red at the
The R&S
NGU switches from voltage priority mode to current priority mode.
Current priority mode