Remote control commands
R&S
®
NGU
164
User Manual 1179.2531.02 ─ 03
HOLD
If the arbitrary function is finished, the last arbitrary point of the
user-defined arbitrary list is held.
*RST:
OFF
Example:
See
Example "Configuring an arbitrary sequence"
ARBitrary:CLEar
Clears the current arbitrary table data for the selected channel.
Example:
See
Example "Configuring an arbitrary sequence"
Usage:
Event
ARBitrary:DATA
<data>
Sets or queries the arbitrary points for the previous selected channel. Max. 4096 arbi-
trary points can be defined. The dwell time between 2 arbitrary points is specified from
1 ms to 20 days.
Depending on the set mode (VPM or CPM) and device model, additional negative cur-
rent or voltage is available for setting.
Parameters:
<data>
voltage1, current1, time1, interpolation mode1, voltage2, cur-
rent2, time2, interpolation mode2...
Voltage and current settings depending on the instrument type
and priority mode (VPM, CPM).
If the interpolation mode is sets to 1, it indicates that the mode is
activated. If the interpolation mode is sets to 0, it indicates that
the mode is not activated.
Example:
ARB:PRI:MODE CPM
Set priority mode to CPM.
ARB:DATA 10,1,0.5,0
Defines one arbitrary point with: voltage1 = 10 V and current1 =
1 A, time1 = 500 ms and interpolation mode1 = 0 (disabled).
ARB:DATA?
-> 10.000, 1.000, 0.50, 0
Returns defined arbitrary points for model R&S NGU201 at CPM
priority mode.
Advanced operating commands