Press the
F1
button again and turn the
M
knob to choose
PK-PK
..
(8)
Read the peak-to-peak voltages of Channel 1 and Channel 2 from the bottom left
of the screen (See
Fig.6-2
).
(9)
Calculate the amplifier gain with the following formulas.
Gain = Output Signal / Input signal
Gain (db) = 20×log (gain)
Fig.6-2
Waveform of Gain Measurement
Example 3: Capture the Single Signal
The digital storage oscilloscope takes the lead in providing the convenience capturing
of such non-periodic signals as pulse and burr, etc. If you intent to capture a single
signal, you cannot set the trigger level and the trigger edge unless you have particular
prior knowledge of this signal. For example, if the pulse is the logic signal of a TTL
level, the trigger level should be set to 2 volts and the trigger edge be set as the rising
edge trigger. If it is uncertain as to the signal, you can make an observation of it in
advance under the automatic or ordinary mode to determine the trigger level and the
trigger edge.
The operation steps are as follows:
(1)
Set the probe menu attenuation coefficient to 10X and that of the switch in the
probe to 10X
(see “How to Set the Probe Attenuation Coefficient” on P14).
(2)
Adjust the
VOLTS/DIV
and
SEC/DIV
knobs to set up a proper vertical and
horizontal ranges for the signal to be observed.