User Manual of Smart Digital Storage Oscilloscope
72
(12)
Press the
F4
button, the
peak-to-peak voltage measurement will be added, finish
settings of channel 2.
Then, the period, frequency, mean and peak-to-peak voltage will be displayed at the
bottom left of the screen and change periodically (see
Fig.6-1
).
Fig.6-1
Waveform of Automation Measurement
Example 2:
Gain of the Amplifier in the Metering Circuit
Set the probe menu attenuation coefficient as
10X
and that of the switch in the probe
as
10X
(see “How to Set the Probe Attenuation Coefficient” on P14).
Connect the oscilloscope CH1 channel with the circuit signal input end and the CH2
channel to the output end.
Operation Steps:
(1)
Push down the
Autoset
button and the oscilloscope will automatically adjust the
waveforms of the two channels into the proper display state.
(2)
Push down the
Measure
button to show the Measure menu.
(3)
Press the
H1
button.
(4)
Press the
F2
button and choose
CH1
.
(5)
Press the
F1
button and turn the
M
knob to choose
PK-PK
..
(6)
Press the
F2
button and choose
CH2
.