User Manual of Smart Digital Storage Oscilloscope
13
How to implement the Function Inspection
Make a fast function check to verify the normal operation of the instrument, according
to the following steps:
1. Connect the power cord to a power source. Turn on the Power Switch
Button
―
○
on the left side (make sure the “
―
” side is pressed down). Then,
push down the button of the “
” signal on the top.
The instrument carries out all self-check items and shows the Boot Logo. Press the
“Utility”
button, then, press
H1
button to get access to the “
Function
” menu. Turn
the
M
knob to select
Adjust
and press
H3
button to select “
Default
”. The default
attenuation coefficient set value of the probe in the menu is 10X.
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope
with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten
the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe
compensator.
3.
Press the “Autoset” Button.
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in
several seconds (see Fig.
4-7
).
Fig.4-7
Auto set
Check CH2 by repeating Step 2 and Step 3.