Centrox Instruction Manual
49
AirSep Corporation
Problem
Probable Cause
Solution
Oxygen concentrator does
not cycle. No green power
light on the PSA module.
Unit not plugged in to the
power supply.
Plug in unit to the electrical
outlet.
No power supply to the wall
outlet.
Replace fuse or reset
breaker.
Blown 15A fuse.
Replace fuse.
Oxygen concentrator does
not cycle, but green power
light is ON on the PSA
module. AUTO/MANUAL
switch is set to MANUAL
position, and blue light is off.
ON/OFF switch is off.
Set ON/OFF switch to ON
position.
Defective ON/OFF switch.
Replace ON/OFF switch.
Blown 1A fuse.
Replace fuse.
Defective AUTO/MANUAL
switch.
Replace AUTO/MANUAL
switch.
Defective power wire to
circuit board.
Repair or replace wire.
Low voltage condition.
Call electric company.
Defective transformer.
Replace transformer.
Defective circuit board.
Replace circuit board.
Содержание Centrox
Страница 1: ...INSTRUCTION MANUAL Centrox PSA Oxygen Concentrator...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 26: ......
Страница 32: ......
Страница 46: ......
Страница 60: ...60 Centrox Instruction Manual AirSep Corporation A 2 DRAWINGS AND SCHEMATICS Figure A 1 Flow Schematic...
Страница 61: ...Centrox Instruction Manual 61 AirSep Corporation Figure A 2 Wiring Schematic 120 V 60 Hz PSA Module...
Страница 62: ...62 Centrox Instruction Manual AirSep Corporation Figure A 3 Wiring Schematic 240 V 50 Hz PSA Module...
Страница 63: ...Centrox Instruction Manual 63 AirSep Corporation Figure A 4 Wiring Schematic 120 V 60 Hz Compressor Module...
Страница 64: ...64 Centrox Instruction Manual AirSep Corporation Figure A 5 Wiring Schematic 240 V 50 Hz Compressor Module...
Страница 68: ......
Страница 71: ......