pNeuton Ventilator
2 - 3
The
pNeuton
Ventilator is MRI Conditional up to
3 T (see Notes section below). While tests show
that the ventilator functions at the bore of the MRI
unit, Airon Corporation does not recommend that the
ventilator be clinically used at or within the bore of MRI
scanners. A
minimum
proximity of 12 inches (0.3 meter)
from the bore should be used. In addition, safe MRI
practice calls for all devices used in the proximity of a MRI
scanner, including
pNeuton,
be anchored to prevent
inadvertent movement.
The Low Gas Supply Alarm will occur if the driving gas
supply drops below safe levels (30 psi, 200 kPa). The
alarm activates as long as driving gas is available or until
supply pressure returns to normal. The alarm will only
activate for a very short period of time if the gas supply
abruptly ceases as can happen if the supply gas becomes
disconnected. Always insure that the supply gas is secure
and operating at the proper pressure.
Cautions
DO NOT
attempt to service the unit. Service may
only be performed by Airon Corporation authorized
engineers. The Preventative Maintenance program
requires a general service and calibration every two years.
Only original manufacturer parts and accessories should
be used.
Any attempts to modify the hardware of this device without
the express written approval of Airon Corporation will void
all warranties and liabilities.
Do not immerse the
pNeuton
Ventilator or allow any liquid
to enter the case or the inlet filter. Clean as directed in
Section 9, Cleaning and Maintenance.
Содержание pNeuton A
Страница 1: ...pNeuton Transport Ventilator Model A Operators Manual 97011 CD A 005 Rev J Released 1 Feb 2019...
Страница 6: ...pNeuton Ventilator 1 2...
Страница 14: ...pNeuton Ventilator 2 8...
Страница 22: ...pNeuton Ventilator 3 8...
Страница 32: ...pNeuton Ventilator 5 4...
Страница 36: ...pNeuton Ventilator 6 4...
Страница 58: ...pNeuton Ventilator 10 4...