Section 9:
Cleaning and Maintenance
Cleaning the Ventilator
Use only mild detergent or disinfectant and water with
a soft cloth.
Do not immerse the ventilator in water.
Do not attempt to sterilize the ventilator with autoclave
or ethylene oxide. Severe damage to the ventilator
may occur.
Cleaning / Disinfecting the Patient Circuit
The Airon patient circuit is a disposable, single use device.
This circuit must not be cleaned, disinfected or reused.
See Single-Use only Medical Device information, page 5-3.
Routine Maintenance
Airon Corporation recommends that an Operational
Verification Test (see Section 4) be performed at initial
installation and prior to use on each patient. Institution’s
standards may require additional biomedical surveillance.
No additional routine maintenance is required.
Factory Preventative Maintenance
Ventilator service is required every 2 years to ensure
continuous safety and reliability of the ventilator.
Ventilator service includes:
o
Replacement of internal filters
o
Replacement of internal materials subject to wear
o
Reconditioning of the enclosure
o
Complete calibration
Содержание pNeuton A
Страница 1: ...pNeuton Transport Ventilator Model A Operators Manual 97011 CD A 005 Rev J Released 1 Feb 2019...
Страница 6: ...pNeuton Ventilator 1 2...
Страница 14: ...pNeuton Ventilator 2 8...
Страница 22: ...pNeuton Ventilator 3 8...
Страница 32: ...pNeuton Ventilator 5 4...
Страница 36: ...pNeuton Ventilator 6 4...
Страница 58: ...pNeuton Ventilator 10 4...