![Airon pNeuton A Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/airon/pneuton-a/pneuton-a_operators-manual_2878642056.webp)
pNeuton Ventilator
10 - 2
Precision - breath to breath repeatability of controls
o
Respiratory Rate + 10%
o
Tidal 25 ml
o
Peak Pr 5 cm H
2
O
o
PEEP / CPAP + 2 cm H
2
O
o
F
I
O
2
+ 5%
Specificity - effect of one control on another
o
Respiratory Rate - if tidal volume is constant, + 5%
o
Tidal 5%
o
Peak Pr 5%
o
PEEP / CPAP + 5%
o
F
I
O
2
+ 5%
Internal Compliance 0.1 ml/cm H
2
O
Ventilator Resistance to Flow
o
Inspiratory, 60 L/min: less then 2 cm H
2
O/l/sec
o
Expiratory, 50 L/min: less then 2 cm H
2
O/l/sec
Alarm System
Patient Disconnect
o
Pressure:
-
mandatory breath OFF, less than 5 cm H
2
O
-
mandatory breath ON, less than 15 cm H
2
O
o
Alarm delay: 22 seconds
o
Alarm silence: 60 seconds
Low Gas Supply
o
Input supply pressure: less than 30 psi (2.1 bar)
o
Cannot be silenced