![Airon pNeuton A Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/airon/pneuton-a/pneuton-a_operators-manual_2878642057.webp)
pNeuton Ventilator
10 - 3
Environmental and Physical Characteristics
MRI Conditional. Tested with a scanner up to:
o
Maximum static field strength - 3 T
o
Maximum spatial field gradient - 720 G/cm
Hypobaric (high altitude) compatible up to 15,000 feet
(4,600 meters)
Weight and Size: 6.5 pounds (3 kg),
5.0"H x 10.0"W x 7.3"D (12.7 cm x 25.4 cm x 18.5 cm)
Storage Temperature Range: -20 to 60 °C (-4 to 140
°F), 15 to 95 percent humidity, noncondensing
Operating Temperature Range: -5 to 40 °C (23 to 104
°F), 15 to 95 percent humidity, noncondensing
Power Sources
Driving gas requirement
o
55 psi + 15 psi (380 kPa + 100 kPa)
o
100% oxygen. Do not use the ventilator with other
types of gases.
o
The gas supply must be capable of delivering at
least 40 liters per minute at 55 psi. If input
pressure drops less than 30 psi due to insufficient
gas flow, the ventilator will alarm and begin to
malfunction.
NOTE
: Driving gas consumption at 10 L/min minute volume;
o
PEEP / CPAP off: 4 L/min
o
PEEP / CPAP on, 65%: 9 L/min
o
PEEP / CPAP on, 100%: 15 L/min