pNeuton Ventilator
4 - 1
Section 4: Operating Instructions
Ventilator Set-up
The following equipment is needed:
1.
pNeuton
Ventilator with breathing circuit (see
Section 5 for a list of compatible circuits.)
2. Test lung, (1 Liter rigid wall, Airon Part # 21002
suggested)
3. Spirometer
4. Watch
When ready:
1. Attach breathing circuit to ventilator as described in
Section 5.
2. Attach the test lung to the patient side of the
breathing circuit.
3. Set the controls as follows:
a.
Mandatory Breath
control to On
b.
% Oxygen
to 65%
c.
PEEP / CPAP
to Off
d.
Peak Pressure
to 50 cm H
2
O
e.
Tidal volume
to 700 ml
f.
Respiratory Rate
to 12 bpm
4. Attach Oxygen Input on rear panel of the ventilator
to a high pressure oxygen source and turn on the
oxygen.
NOTE:
The ventilator will begin operation at the above
settings when the oxygen is turned on. The alarm will
sound. You may press the “Reset / Silence” button to
silence the alarm or wait for the unit to begin ventilating.