background image

Section 5:  Integrated Gas Sampling

Airmetrics

39

1)  Place the Overlap Jumper on the pins.

2)  Plug ba g 1 solen oid into So lenoid O utput Co nnector 2 ; plug bag  2 soleno id
into Solenoid Output Connector 4.

3) Set the timer: 

a)  Set timer Program 1 to turn on at 10:00 am and off at 3:00 pm. During

this program, bag 1 will be filling.

b)  Set timer Program 2 to turn on at 3:01 am and off at 6:00 pm. During

this program, both solenoids will be active and both bags will be filling
simultaneously.

c)  Set timer Program 3 to turn on at 6:01 pm and off at 11:00 pm. During

this programmed step, only bag 2 will be filling.

A

DJUSTING 

P

ULSE 

F

REQUENCY AND 

D

URATION

 The rate at which the bags are filled is set by using a tunable intervalometer or
pulse circuit which can be adjusted both for frequency (continuously on to 1 pulse
in 15 seco nds) and  for duration  (50 ms to  750 m s). The pu lse freque ncy is
controlled by the 

Intervalometer Frequency Adjustment

, while the duration of

each pulse is set by the 

Pulse Du ration Adju stment.

 The 

Power o n/off

 (sw-5)

enables gas sampling mode.

Adjusting the pulse circuit using the Intervalometer Frequency Adjustment and the
Pulse D uration A djustme nt is accom plished thr ough trial a nd error. A  test period in
the labora tory is there fore requir ed before  the sam pler can b e mov ed to the field
site. The o bject is to ac hieve a co mbine d pulse d uration an d interval th at will
integrate a sample of air over the programmed period of time. At the end of the
programmed period the bags should be 80-90% filled. That is, the bag should not
be tightly filled, since there would be no way of knowing at what point the bag
became filled.

The pu lse duration  and frequ ency co ntrols can  be adjuste d to suit the re quirem ents
of the task at hand. For example; one can take many small samples or a few large
samples ov er the same p eriod of time dep ending on the n eeds of the opera tor.

Airmetrics MiniVol Users Guide

40

Airmetrics

Pulse Interval Adjustment

The pulse interval (off time) of the circuit is adjustable over a range of 0-15
seconds by a 16-position rotary switch. This switch is located just to the right of
the solenoid valve output connectors. Switch positions are marked clockwise 0-9,
and continue A-F. The interval between pulses increases in 1-second increments as
the switch is rotated in a clockwise direction. Position "0" enables continuous
pulsing. Position "1" corresponds to a minimum delay time of one pulse per
second, and position "F" indicates the maximum delay of 15 seconds between

pulses. The fine adjustment potentiometer next to this control is for calibrating the
one-second  interval (fully counterclock wise - decrease , fully clockwise - increa se).

Pulse Duration Adjustment

The duration (on time) of each pulse is adjustable over a range of 50-750
milliseconds. This adjustment is made by summing the calibrated interval values of
DIP switches 1-4 on the 5-position DIP switch located in the upper right corner of
the auxiliary board. Each switch has an assigned "On Time" value, and the on time
for each pulse is de termined by th e sum of the v alues of switches  in the "off"
position. The "On Time" v alues for each switch are as follows:

SW-1 =   50 ms
SW-2 = 100 ms
SW-3 = 200 ms
SW-4 = 400 ms

For Example: If SW-1, and SW-4 are in the "off" position, the solenoid on time
for each pulse w ould be 450 m illiseconds. (50 +  400 ms =  450 ms). 

The D OW N positio n is "on" fo r these sw itches. Th e fine adju stment n ext to this
control adds an extra 50 ms to the total pulse time (fully counterclockwise - adds
50 ms, fully clock wise - adds no  time).

Calculate an approximate pulse duration and frequency that would fill the 6 liter
Tedlar® bag over the desired collection period.

Example:

 Suppose you wish to fill a 6 liter bag over a period of 4 hours. The first

step wo uld be to d etermine  the amo unt of air pu mped  during a s ingle pulse . This
can be done by replacing the bag at the mini quick-connect with a bubble flow
meter, and starting the pump in the gas sampling mode. If 1 cc of air was pumped

Содержание MiniVol

Страница 1: ...ing History Manua l Version 1 0 O ctober 1992 ap plies to Sampler M odels 3 x Manua l Version 1 0a N ovembe r 1992 applies to Sa mpler M odels 3 x Manua l Version 1 1 Jun e 1993 applies to S ampler M...

Страница 2: ...e Button 18 Airmetrics MiniVol Users Guide ii Airmetrics Pulse Interval Adjustment 18 Pulse Duration Adjustment 18 Power O n Off 19 Pulse Indicator 19 Overlap Jumper 19 4 PARTICULATE MATTER SAMPLING 2...

Страница 3: ...QUICK REFERENCE 79 D WARRANTY POLICY 83 E REPLACING DAMAG ED DEFECTIVE COMPONENTS 85 F PARTS LIST 87 G FORMS and SCHEMATICS 105 List of Figures Figure 2 1 Attaching Battery Pack 7 Figure 2 2 Programm...

Страница 4: ...circuits Airmetrics MiniVol Users Guide 2 Airmetrics A low battery circu it automatically shuts the sampler down should the rechargeable lea d acid battery fail to supply sufficient voltage below 10 3...

Страница 5: ...proven to be an excellent indicator of absolute ambient PM10 concentrations Although the method used by portable PM10 sampling does not wholly conform or comply with the reference method the data col...

Страница 6: ...th local plug configura tion to connect the tran sformer to a wall socket for battery recharge 3 The green LED on the top of the battery will light indicating that the battery is being ch arged W hen...

Страница 7: ...ontrolled by prog ramme d on off sequences As the ON AUTO OFF button is pressed a bar at the lower edge of the LCD display moves horizontally over the words On Auto and Off which are printed on the ti...

Страница 8: ...h e on time In this way sampling may start on one day and end on the next day Airmetrics MiniVol Users Guide 10 Airmetrics L 5 Press the PR OG button again 2ON appears on the display to indicate that...

Страница 9: ...t will not accurately reflect the amount of air passing through the filter The sampler will be measuring air passing through the filter plus whatever air may be entering through the leak Verify that a...

Страница 10: ...tes the flow rate of air through the system in liters minute The flow rate is adjusted using the Flow Rate Adjustment The flow meter rea dings m ust be take n from th e center of the ball FLOW RATE AD...

Страница 11: ...low meter Airmetrics MiniVol Users Guide 16 Airmetrics L 3 Set the Lo w Flow Cutoff A djustme nt to the zero position rotated fully counterclockw ise 4 Adjust the Low Flow Zero Adjustment until the Lo...

Страница 12: ...en the sampler is switched to the O n mode either m anually o r by programm ed opera tion Wh en a solen oid output is ac tive and the samp ler is in the O n mode a solen oid valve plugged onto the act...

Страница 13: ...ockwise adds 50 ms fully clock wise adds no time POWER ON OFF Switch 5 on this DIP switch is the on off power switch for the auxiliary board and must be set to the o n position for any fun ctions to b...

Страница 14: ...ility pole parking meter fence po st etc See Figure 4 1 Available separately from Airmetrics is a Y Bracket Assembly which attaches to poles and provides a mast for the mounting cradle Also sho wn in...

Страница 15: ...y malfunctions before proceeding Perform a single point flow rate check using a soap bubble meter or other flow measuring device of known accuracy and compare to the curve established during calibrati...

Страница 16: ...e flo wme ter or the pu lse dam pener is replaced Preseparator Filter Holder Assembly Depending on the required particle size separation the configuration of the Preseparator changes The attached Filt...

Страница 17: ...Section 4 Particulate Matter Sampling Airmetrics 27...

Страница 18: ...p of the bale assembly bar and remove the bale assembly Airmetrics MiniVol Users Guide 30 Airmetrics 6 Lift the pump and timer assembly out by the 6 diameter top cap and rest it on the edge of the sam...

Страница 19: ...e top cap 5 Check the sampler face plate for any error conditions If an error conditions exists refer to the Error C onditions section at the end of this chap ter 6 Verify correct time and day of week...

Страница 20: ...n dix D Low Flow Indicator ON Should the Low Flow Indicator be found ON at the end of sam pling period first check the Elapsed Time Totalizer to determine the length of time the sampler ran before shu...

Страница 21: ...e Mount the M iniVol cra dle onto a standard 1 1 4 inch antenna mast or c omparable metal tubing which itself is strapped securely to another supporting structure utility pole parking m eter fence pos...

Страница 22: ...d into Solenoid Output Connector 3 2 Set the timer Airmetrics MiniVol Users Guide 38 Airmetrics a Set timer Program 1 to turn on at 1 00 am and off at 7 00 am During this program neither bags will be...

Страница 23: ...rmetrics MiniVol Users Guide 40 Airmetrics Pulse Interval Adjustment The pulse interval off time of the circuit is adjustable over a range of 0 15 seconds by a 16 position rotary switch This switch is...

Страница 24: ...A F where zero is continuous pulsing and F is one pulse eve ry 15 seconds 7 Turn on the valve sequencer board by moving the P ower O N OF F Switch to the UP position Note that the other four dip switc...

Страница 25: ...unobstructed area at least 30 cm from any obstacle to air flow 2 With the sample r on a firm level surfac e unscrew either c ap of bale assembly bar and remove bale assembly 3 Detach canister tubing...

Страница 26: ...4 4 2 With the sample r on a firm level surfac e unscrew either c ap of the b ale assembly bar and remove bar assembly Detach canister tubing at the mini quick connect fittings and remove canisters 3...

Страница 27: ...outlet of the flowmeter The Flow Control System is temperature compensated for reasonable changes in ambient temperature and pressure MINIATURE BRUSHLESS D C DOUBLE ACTING DIAPHRAGM PUMP The Double Ac...

Страница 28: ...e charg er is designed to quick charge th e battery and then switch to a mainte nance mode to avoid an overcharge condition The charger can be left on for an indefinite period without damaging the bat...

Страница 29: ...actor cleaned and grea sed at regu lar intervals i e every seventh sam ple to start bu t if heavy loadings are observed on the target disk as often as appropriate 1 Unm ate the preseparator sections f...

Страница 30: ...by dirt they will become worn and n eed replacemen t The side of the pump on which the valves are worn or dirty is easily determined by pinching the inlet tube s leading to the pum p first one side a...

Страница 31: ...MiniVol Users Guide 56 Airmetrics sampler to charge the batteries 2 Disassemble the battery case using a Phillips screwdriver and connect the battery charger board to another battery If the same cond...

Страница 32: ...rately adjusted using the Flow R ate Adju stment The pump valves and diaphragms are dirty or worn and need cleaning or replacing see Section 7 Maintenance Cleaning Replacing Pump V alves and D iaphrag...

Страница 33: ...r stream during the calibration procedu re Figure A 1 shows the results of a typical MiniVol calibration The columns in the table in Figure A 1 are defined as Qind Flowrate as ind icated by the rotam...

Страница 34: ...Appendix A Calculating Flow Rate Airmetrics 61...

Страница 35: ...pt bvol of the sampler calibration are obtained from the calibration sheet For S ampler 231 8 mstd 1 0011 and bstd 0 0826 The units of Qstd are liters minu te 2 Calculate the volume of air that passed...

Страница 36: ...ter by the volume of air that passed throug h the filter or where PM act PM concen tration in g rams pe r cubic m eter actual PM std PM concentration in grams per cubic meter standard MPM M ass of par...

Страница 37: ...degrees and range over normal room temperatures Airmetrics MiniVol Users Guide 68 Airmetrics 7 Staticmaster ionizing units One is centered in balance chamber another mounted c onveniently on stand ne...

Страница 38: ...r month period of the series of substitute weights A total of 75 meas urements m ade on six different w eights resulted in a rang e of 1 Airmetrics MiniVol Users Guide 70 Airmetrics microgram on the 2...

Страница 39: ...ri dishes These filters are to be weighed at the beginning con trol filter masses C1ca 2ca 3ca and end control filter masses C1cb 2cb 3cb of each weigh ing session for any filters from the same lo t T...

Страница 40: ...tings Field sampling reports or data sheets are sent with each filter or set of filters Figure 3 7 Filter heads or Petrislides received from the field are first checked against the accompanying field...

Страница 41: ...pleted the field sampling report may be filed Airmetrics MiniVol Users Guide 76 Airmetrics 6 The computer program used to calculate particulate matter concentrations from portable saturation samplers...

Страница 42: ...duce most of the variability in filter weights close control must be maintained on these factors A ny chan ge of 1 in humidity or of on e degree celsius in temperature during a weighing session necess...

Страница 43: ...his recording the actua l filter number 9 Remove the preseparator filter holder assembly at the quick connect and place Airmetrics MiniVol Users Guide 80 Airmetrics it in clean plastic bag for transpo...

Страница 44: ...CD 5 Record elapsed tim e as show n on the E lapsed T ime To talizer wh ich should match the programmed time 6 Check and record the fullness of the bag s adjusting the pulse interval or duration if ne...

Страница 45: ...warranty does not apply if the product has been damaged by accident or misuse or as the result of service or modification by other than an Airmetrics technician No other express warranty is given The...

Страница 46: ...541 434 9695 Obtain authorization to return the defective or damaged compo nents Package the item s carefully to prevent further damage Identify the item s being returned on a clearly marked packing l...

Страница 47: ...List Airmetrics 87 F PARTS LIST The following Figures and Tables identify the components and parts of the MiniVol sampler Airmetrics MiniVol Users Guide 88 Airmetrics Figure F 1 Main Circuit Board Par...

Страница 48: ...lastic Y connector 3 16 101 014 Inlet Hose Barb Fitting 1 4 MNPT 3 16 ID H ose 101 015 Flat Washer s s 9 16 SAE 101 003 Flowmeter 0 10 LPM 102 000 Pump Dbl Diaphragm 8 12 V 102 001 Pump Hose B arb Fit...

Страница 49: ...Valve D river Board 103 001 Valve Drive r Circuit Board 103 002 Solenoid Valve 103 003 Solenoid Output Orifice Fitting 10 32 1 8 ID Ho se 104 001 Panel M ount M ini Quick Conne ct male 104 011 Female...

Страница 50: ...lon Bolt 1 2 13 3 4 104 003 Male Quick Connect Inlet Fitting 104 004 Battery Pack Mounting Latch 104 005 Banana Plug 104 006 Nylon Hex Nut 1 2 13 105 000 Interconnect Bo ard 105 002 Power C able 105 0...

Страница 51: ...ttom 206 002 03 O Ring for Top Teflon 206 002 04 O Ring for Bottom Viton 206 003 Quick D isconnect 1 4 m pt female 206 004 Preseparator Tube Grey 201 000 PVC Rain Hat 202 000 QC Impactor PM10 QC C omp...

Страница 52: ...ics MiniVol Users Guide 98 Airmetrics Battery Pack Assembly see Figure F 5 P N Description 300 000 Battery Pack As sembly 301 000 Battery Charg er Board 301 004 Charging Jack 302 000 Battery 12V 12AH...

Страница 53: ...niVol Users Guide 100 Airmetrics Integrated Gas Sampling Bag Assembly see Figure F 6 P N Description 400 000 Com plete Bag Module 400 001 Tedlar Bag 5 liter 400 003 Mini Quick Connect Hose Fitting 400...

Страница 54: ...Appendix F Parts List Airmetrics 101...

Страница 55: ...ide 104 Airmetrics G FORMS and SCHEMATICS The forms shown in this appendix are samples of the Field Data Sheets that may be used when performing PM sampling Figure G 1 and Integrated Gas sampling Figu...

Страница 56: ...Appendix G Forms and Schematics Airmetrics 105 This page was intentionally left blank...

Отзывы:

Похожие инструкции для MiniVol