22
T R O U B L E S H O O T I N G
Trouble
Possible Cause
Corrective Action
7. Image appears
skewed on monitor.
• PSP was fed skewed.
• When inserting PSP into
feed slot, be sure to “feel”
for resistance of light seal
brush, align PSP, and then
push down uniformly on
top edge of PSP.
8. Image contains ghost
images or shadows.
• PSP was not completely
erased prior to use.
• Imaging Plate was exposed
with the back of the IP facing
the tubehead.
• PSP has been stored too
long in X-ray cassette.
• Partial erasure of the image
due to exposure to light
during handling of the PSP
• Make ScanX is operating
with eraser turned on
(blue LED indicator above
the ERASER switch is
illuminated).
• Make sure the plates are
inserted properly into the
cassette with the proper ori-
entation to the X-ray source.
• Do not store PSPs in
cassettes for more than
24 hours.
• Do not leave exposed
PSPs in well lit areas.
Transfer PSPs from their
X-ray cassette to the
ScanX within one hour of
exposure.
9. Image shows
artifacts or white or
black lines.
• The PSP surface is not clean
and has dirt, stains or
scratches on it.
• ScanX plate transport path
may contain an obstruction,
debris or dust.
• Clean the PSP. If the PSP is
scratched or stained, do
not reuse.
• Clean transport path using
a ScanX Cleaning Sheet.