T E C H N I C A L D A T A
Electrical Requirements:
Supply Voltage:
100 to /- 10%, 50/60 Hz
Supply Current:
1.2 A Maximum
Line Cord:
North American style 10 foot long
Hospital Grade line cord, P/N 73096.
Power Supply:
24 Volt Power Supply, P/N B7095 provided
Physical Properties:
Dimensions
Depth:
15.4 in.
Width:
15.0 in.
Height:
27.8 in.
Weight:
50 lbs.
Environmental Conditions:
Unit in Operation
Temperature:
50°F to 105°F (10°C to 40°C)
Humidity:
5% to 95% RH
Heat emission:
<90W
Storage and Transport
Temperature:
-21°F to 130°F (-29°C to 54°C)
Humidity:
5% - 95% (Non-condensing)
Note:
Resolution of the ScanX is dependent on operating mode and specific
imaging plate type used.
Resolution Range (lp/mm):
Horizontal:
5.0
Vertical:
5.0
Compliance Data:
Laser Class:
Class I Laser Product (21 CFR 1040.10)
Class 1 Laser Product (IEC 60825-1)
Laser Product Report
Accession Number:
0212282-00
Installation Category:
I
Pollution Degree:
2
User Replacement Items:
Power Supply, Line Cord, PSPs
and Computer Connector Cable
IEC601-1 Classification:
Class 1, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic
mixture(s). Protection against ingress of liquids -Ordinary
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can
occur. Do not use the equipment in close conjunction with sensitive devices, or
devices creating high electromagnetic disturbances.