T R O U B L E S H O O T I N G
Trouble
Possible Cause
Corrective Action
1. No power/
No green light on
membrane switch
• Not plugged in.
• No power at Mains
Outlet
• Defective power supply
• Check the line cord
connection is firmly
plugged in.
• Make sure outlet is
grounded and has power.
• Call your authorized dealer.
2. Green, Blue or
Yellow indicator light
does not work.
• Defective light or circuitry.
• Call your authorized dealer.
3. Imaging Software
does not recognize
the ScanX when
selected.
• Inadequate Computer
System.
• The ScanX has not been
turned on.
• The computer connection
cable is loose or defective.
• The computer does not
recognize that the ScanX is
connected.
• ScanX hardware problem.
• Verify Computer System
requirements (Page 10).
• Make sure that the READY
switch is set to ON and
the green indicator light is
illuminated.
• Reconnect the cable.
Check for tightness.
Replace if necessary.
• Verify that the Setup
program was correctly
installed (Page 14)
.
• Call your authorized dealer.
4. Plate does not
scan properly.
• The PSP was not pushed
far enough into the ScanX.
• Fully feed the PSP into the
Inlet Ring.
5. No image appears
after scanning.
IIm
mp
po
orrtta
an
ntt::
Do not allow the
PSP to be exposed
to light between
taking an X-ray
and scanning with
the ScanX.
• The PSP fed backwards
(printed side out ScanX).
• The PSP was erased prior to
scanning.
• Hardware failure.
• X-ray source failed
or low exposure.
• Quickly refeed the plate
with the printed side out. If
a substandard image
results, retake image.
• Feed the PSPs into the
scanner immediately and
quickly after removal from
the X-ray cassette.
• Call your authorized dealer.
• Call your authorized dealer.
6.
Image is too dark.
• PSP has been overexposed
• Use software to adjust
brightness. If this is not
possible, retake image
with proper (lower)
exposure.