T R O U B L E S H O O T I N G

Trouble

Possible Cause

Corrective Action

1. No power/

No green light on
membrane switch

• Not plugged in.

• No power at Mains

Outlet

• Defective power supply

• Check the line cord

connection is firmly
plugged in.

• Make sure outlet is

grounded and has power.

• Call your authorized dealer.

2. Green, Blue or

Yellow indicator light
does not work.

• Defective light or circuitry.

• Call your authorized dealer.

3. Imaging Software

does not recognize
the ScanX when
selected.

• Inadequate Computer

System.

• The ScanX has not been

turned on.

• The computer connection

cable is loose or defective.

• The computer does not

recognize that the ScanX is
connected.

• ScanX hardware problem. 

• Verify Computer System

requirements (Page 10).

• Make sure that the READY

switch is set to ON and
the green indicator light is
illuminated.

• Reconnect the cable.

Check for tightness.
Replace if necessary.

• Verify that the Setup

program was correctly
installed (Page 14)

.

• Call your authorized dealer.

4. Plate does not

scan properly.

• The PSP was not pushed

far enough into the ScanX.

• Fully feed the PSP into the

Inlet Ring.

5. No image appears

after scanning.

IIm

mp

po

orrtta

an

ntt::  

Do not allow the
PSP to be exposed
to light between
taking an X-ray
and scanning with
the ScanX.

• The PSP fed backwards

(printed side out ScanX).

• The PSP was erased prior to

scanning.

• Hardware failure.

• X-ray source failed

or low exposure.

• Quickly refeed the plate

with the printed side out. If
a substandard image
results, retake image.

• Feed the PSPs into the

scanner immediately and
quickly after removal from
the X-ray cassette.

• Call your authorized dealer.

• Call your authorized dealer.

6.

Image is too dark.

• PSP has been overexposed

• Use software to adjust

brightness. If this is not
possible, retake image
with proper (lower)
exposure.

Содержание ScanX B7300

Страница 1: ...Instruction Manual Part Numbers B7300 ...

Страница 2: ...art Number Imaging Plate Intensifier Screen Warning Instructions 73020 Phosphor Storage Plate Instruction Sheet 73474 Section Page Congratulations 3 Warranty 3 On Line Warranty Registration 3 Safety Notice 4 Important Information 7 Purpose of this Manual 8 System Description 8 Unpacking and Inspection 9 Computer System Requirements 10 Controls and Indicators 11 Technical Data 12 Abbreviations 13 S...

Страница 3: ... models of the ScanX are designed solely for use in an o of ff fi ic ce e e en nv vi ir ro on nm me en nt t and this warranty is not appli cable to other applications Any item returned to our factory during the warranty period through an a au ut th ho or ri iz ze ed d d de ea al le er r will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have b...

Страница 4: ...e or explosion DANGER AVOID DIRECT EXPOSURE TO BEAM LASER RADIATION WHEN OPEN CAUTION CLASS 1 LASER PRODUCT CLASS 3B LASER RADIATION WHEN OPEN AVOID EXPOSURE TO BEAM CLASS IIIb LASER PRODUCT COVER REMOVED MAKES THIS DEVICE A Alerts users to important Operating and Maintenance instructions Read carefully to avoid any problems Warns users that uninsulated voltage within the unit may be of sufficient...

Страница 5: ...tools It should be opened and serviced only by an authorized dealer service technician Failure to heed this warning may result in equipment damage or personal injury and w wi il ll l v vo oi id d a an ny y a an nd d a al ll l w wa ar rr ra an nt ti ie es s Contact your authorized dealer for service information Electrical Safety Notes K The line cord is the main power disconnect device K Use only t...

Страница 6: ... this device to individuals other than licensed professionals Use of this device other than as described in this manual may result in injury The ScanX contains a laser and is a Class 1 Class 1 IEC Laser Product Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure The laser is on only during an active scan Only a tr...

Страница 7: ... are fully dry again K The manufacturer regard themselves as being responsible for the equip ment with regard to safety reliability and proper functioning only if assem bly resetting changes or modifications and repairs have been carried out by an authorized dealer and if the equipment is used in conformity with the instructions contained in this manual K The device conforms to the relevant safety...

Страница 8: ...rrect usage The manufacturer is not to be held liable for any damage caused as a result of incorrect usage The operator bears all risks This manual provides the information necessary for the setup operation and routine care and maintenance of the ScanX 12 Digital Imaging Systems with In Line Erase P N B7300 This manual is not to be used as a replacement for training in radiography For information ...

Страница 9: ...ned to Air Techniques Inc ALLPRO Imaging for a full refund of the fee using a Return Authorization RA number or retain without refund of the fee to protect the ScanX during transport or future shipping Retain the over packag ing to return the shipping case Included System Components The ScanX consists of the model and accessory kit as listed below See TECHNICAL DATA for ratings and identification ...

Страница 10: ...ecessary to operate the ScanX are listed below Items listed with an asterisk are required Operating System Windows 2000 Professional for an Intel 32 bit processor with Service Pack 4 or later or Windows XP Professional for an Intel 32 bit processor with Service Pack 1and the KB B22603 update or later USB Port Must be USB 1 1 or 2 0 or later CPU 2 8 GHz Pentium IV RAM 1 GB Hard Drive 200MB availabl...

Страница 11: ...dby Green LED indicator K Ke ey yp pa ad d C Co on nt tr ro ol l P Pa an ne el l READY Switch Toggles between the Standby and Ready mode as follows 1 Press to switch from the Standby mode to the Ready mode 2 Press and hold down for at least 2 seconds to switch to the Standby mode from the Ready mode Ready Standby Green LED indicator Illuminates green to indicate that the ScanX is Ready for operati...

Страница 12: ...ing mode and specific imaging plate type used Resolution Range lp mm Horizontal 5 0 Vertical 5 0 Compliance Data Laser Class Class I Laser Product 21 CFR 1040 10 Class 1 Laser Product IEC 60825 1 Laser Product Report Accession Number 0212282 00 Installation Category I Pollution Degree 2 User Replacement Items Power Supply Line Cord PSPs and Computer Connector Cable IEC601 1 Classification Class 1 ...

Страница 13: ...SITE SELECTION A ampere s AC alternating current CD ROM compact disk read only memory CFR Code of Federal Regulations CPU central processing unit your computer cm centimeter D depth GB gigabyte 230 109 bytes GHz Gigahertz 109 of hertz H height Hz hertz cycles per second IEC International Electro tech nical Commission IMS Image Management Software also referred to as Patient Management Software PMS...

Страница 14: ...s installed properly on the com puter 4 Verify that the supplied ScanX Drivers and Utilities Disk which contains the USB drivers was properly installed per instructions above Note Always make sure to use the same USB port whenever re connection of the USB cable is necessary 5 Connect the high speed USB cable between the USB Type B connector located on the ScanX rear panel and the USB Type A connec...

Страница 15: ...ector Cable Power Supply Line Cord Press HERE to turn ON or OFF Green LED Indicator Mains Outlet Line Cord The Line Cord is the Mains disconnect device TO POWER SUPPLY TO COMPUTER USB PORT ALLPRO IMAGING READY ERASER FLASHING OFF ALLPRO IMAGING READY ERASER FLASHING OFF Blue LED Indicator ERASER Switch READY Switch SYSTEM SETUP ...

Страница 16: ...Use a Plate Protector for each plate PSPs may be exposed in cassettes and scanned without Plate Protectors however greater care must be exercised to not expose the PSP to light before scanning and to pre vent the PSP from being scratched or soiled Plate Protection When storing or transferring PSPs use an X ray Cassette for PSPs Plate Protector A correct size Plate Protector should be used when han...

Страница 17: ...that should be subsequently deleted from the imaging software Cleaning Phosphor Storage Plates For the best images PSPs should be handled carefully and kept clean Use the following procedure to clean plates 1 Use lint free 100 cotton gauze not cotton balls Gently wipe the cotton gauze over the dry Plate surface Wipe back and forth and then in a circular motion 2 To clean any remaining stains dampe...

Страница 18: ...image type and resolution 5 Verify that the four Scanner Track indicators illuminate green when the Scanner has been activated indicating that a PSP can be fed into the ScanX Note Only one exposed PSP can be fed into the ScanX at a time The next PSP may be fed only after all four track indicator light LEDs change from yellow to green Scanning and Erasing Plates Scan and erase an PSP in one operati...

Страница 19: ... off When the eraser mode is disabled the ScanX scans the same as when the eraser is enabled except that the PSPs are n no ot t erased after scanning Scan an PSP without erasing the image by performing the following procedures 1 Activate the scanner by performing the procedures on previous page Note Upon activation the ScanX defaults with the eraser mode enabled This must be disabled prior to scan...

Страница 20: ... Transport Over time small debris and dust can accumulate in the plate transport mecha nism causing a loss in image quality and possible damage to the PSPs To ensure optimal performance of the ScanX the plate transport should be cleaned once per week using a new ScanX Cleaning Sheet each time Sample sheets are included with the ScanX and additional sheets can be purchased from your dealer Refer to...

Страница 21: ... light is illuminated Reconnect the cable Check for tightness Replace if necessary Verify that the Setup program was correctly installed Page 14 Call your authorized dealer 4 Plate does not scan properly The PSP was not pushed far enough into the ScanX Fully feed the PSP into the Inlet Ring 5 No image appears after scanning I Im mp po or rt ta an nt t Do not allow the PSP to be exposed to light be...

Страница 22: ...ring handling of the PSP Make ScanX is operating with eraser turned on blue LED indicator above the ERASER switch is illuminated Make sure the plates are inserted properly into the cassette with the proper ori entation to the X ray source Do not store PSPs in cassettes for more than 24 hours Do not leave exposed PSPs in well lit areas Transfer PSPs from their X ray cassette to the ScanX within one...

Страница 23: ...Kit 1 74050 Line Cord 1 73096 24 V Power Supply Adapter 1 B7095 6 Foot Computer Connector Cable USB 1 B3554 15 Foot Computer Connector Cable USB 1 B5026 Medical Phosphor Storage Plates 8 x 10 inches 1 per package 73578 8M 10 x 12 inches 1 per package 73578 10 Plate Protectors 8 x 10 inches 1 per package B2160 8M 10 x 12 inches 1 per package B2160 10 Cassettes 8 x 10 inches 1 per package 74190 8 10...

Страница 24: ...ts for the dental medical and veterinary professional since 1962 Air Techniques and ALLPRO Imaging products are distributed only through authorized dealers Refer to www airtechniques com or www allproimaging com to find a dealer in your area K AccentJ Intraoral Digital X ray Image System K Acclaim USB Only Intraoral Digital Video Camera System K Acclaim Intraoral Digital Video Camera System K AirS...

Отзывы: