Page 15
Air Techniques, Inc.
Equipment Setup
46.00 in.
(1165 mm)
45.25 in.
(1150 mm)
5.1 Installation Procedure Summary
This section provides procedures to install the Provecta HD . Install the Provecta HD by performing the
procedures listed below and provided by the following paragraphs .
1 . X-ray base unit Wall Mounting Options
- Single Post Mount (4 x 4 inch post) installation using mounting plate P/N A6389
- 16 Inch on Center Mount installation using mounting plate P/N A6341
2 . Control unit disassembly for base unit access
3 . Base unit lock slide setup
4 . Base unit with exposure keypad installation
5 . Horizontal arm and scissor arm installation
6 . Tubehead installation
7 . Determine optional remote switch configuration and install as necessary .
(See kit, part number A6320 .)
8 . X-ray control and main power connections
5 Installation